Output factor measurements for a kilovoltage x-ray therapy unit

被引:25
作者
Healy B.J. [1 ]
Gibbs A. [1 ]
Murry R.L. [1 ]
Prunster J.E. [1 ]
Nitschke K.N. [1 ]
机构
[1] Southern Zone Radiation Oncology Service - Mater Centre, Princess Alexandra Hospital, South Brisbane, QLD 4101, Raymond Tce.
来源
Australasian Physics & Engineering Sciences in Medicine | 2005年 / 28卷 / 2期
关键词
Ionisation chambers; Kilovoltage x-rays; Output factors; Water substitutes;
D O I
10.1007/BF03178702
中图分类号
学科分类号
摘要
Output factors at the surface for treatment cones and lead cut-outs have been measured for a Pantak Therapax SXT 150 superficial therapy unit with x-ray beam qualities from 1 to 13 min A1 HVL. A variety of phantom materials and two ionisation chambers were tested for their suitability in output factor and percentage depth dose measurement. Solid water proved a useful water-equivalent phantom material with discrepancies between measurements in water and solid water less than 2.3% for percentage depth dose and less than 0.6% for output factors. Larger measurement discrepancies were found for Plastic Water and Perspex. A PTW Markus chamber was found to compare well with a NE 2532/3 low energy chamber in percentage depth dose measurement, but discrepancies arose between the chambers in output factor measurements, up to 5% for small field sizes. Measurements indicated that the Markus chamber had an energy dependent response in the kilovoltage range, which could account for the discrepancy in output factor measurement. Copyright © 2005 ACPSEM/EA.
引用
收藏
页码:115 / 121
页数:6
相关论文
共 20 条
  • [1] The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 m Al - 4 mm Cu HVL
  • [2] 10 -300 kV generating potential), Phys. Med. Biol., 41, pp. 2605-2625, (1996)
  • [3] Absorbed Dose Determination in External Beam Radiotherapy Technical Report Series No. 398, (2000)
  • [4] AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology, Med. Phys., 28, pp. 868-893, (2001)
  • [5] Allen Li X., Ma C.-M., Salhani D., Relative dosimetry measurement for kilovoltage x-ray units, Kilovoltage X-ray Beam Dosimetry for Radiotherapy and Radiobiology, (1999)
  • [6] Ipe N.E., Rosser K.E., Moretti C.J., Manning J.W., Palmer M.J., Air kerma calibration factors and chamber correction values for PTW soft x-ray, NACP and Roos ionisation chambers at very low x-ray energies, Phys. Med. Biol., 46, pp. 2107-2117, (2001)
  • [7] Jones D.E.A., Raine H.C., Br. J. Radiol., 22, pp. 540-550, (1949)
  • [8] Perrin B.A., Whitehurst P., Cooper P., Hounsell A.R., The measurement of k<sub>ch</sub> factors for application with the IPEMB very low energy dosimetry protocol, Phys. Med. Biol., 46, pp. 1985-1995, (2001)
  • [9] Klevenhagen S.C., The build-up of backscatter in the energy range 1 mm Al to 8 mm Al HVT, Phys. Med. Biol., 27, pp. 1035-1043, (1982)
  • [10] Podgorsak E.B., Gosselin M., Evans M.D.C., Superficial and orthovoltage x-ray beam dosimetry, Med. Phys., 25, pp. 1206-1211, (1998)