Studies on the Biotribological and Biological Behavior of Thermally Oxidized Ti6Al4V for Use in Artificial Cervical Disk

被引:0
作者
Song Wang
Junhui Li
Junzhe Lu
Rajnesh Tyagi
Zhenhua Liao
Pingfa Feng
Weiqiang Liu
机构
[1] Tsinghua University,Department of Mechanical Engineering
[2] Research Institute of Tsinghua University in Shenzhen,Biomechanics and Biotechnology Lab
[3] Henan Luoyang Bonesetting Hospital,The Spine Surgery of Henan Orthopaedic Hospital
[4] Indian Institute of Technology (BHU) Varanasi,Department of Mechanical Engineering
[5] Tsinghua University,State Key Laboratory of Tribology
来源
Journal of Materials Engineering and Performance | 2017年 / 26卷
关键词
artificial cervical disk; biological behavior; simulation; thermal oxidation; Ti6Al4V alloy; wear;
D O I
暂无
中图分类号
学科分类号
摘要
The artificial cervical disk was simplified and considered as a ball-on-socket model with the material configuration of ultra-high molecular weight polyethylene and Ti6Al4V (PE-on-TC4). In order to improve the wear resistance, an optimized thermal oxidation (TO) coating was applied on TC4 component. The long-term wear behavior of the model was assessed in vitro using a wear simulator under 10 million cycles (MC) testing intervals. The biological behavior was investigated by bone marrow-derived mesenchymal stem cells (BMSCs) cell attachment and cell viability/proliferation assays, respectively. The total average wear rate for PE/TC4 pair was found to be 0.81 mg/MC, whereas the same was about 0.96 mg/MC for PE/TO pair. The wear rate of the metal has been neglected in comparison with that of the mating polymer. PE component was found to suffer severe damage characterized by scratches, fatigue cracks and arc-shaped wear grooves on the edge zone of ball. The dominant wear mechanism was abrasion for metal component while the dominant failure mechanism was a mix of plowing, fatigue and plastic deformation for polymer component. TO coating improved the cell attachment property of TC4, and the cell viability results were also quite good. TO coating protected TC4 from being plowed and avoided the release of toxic metal ions. However, this intensified the wear of PE component. Considering the biotribological and biological behavior in totality, TO coating could still be promising when applied in articulation surfaces.
引用
收藏
页码:2271 / 2284
页数:13
相关论文
共 50 条
  • [1] Studies on the Biotribological and Biological Behavior of Thermally Oxidized Ti6Al4V for Use in Artificial Cervical Disk
    Wang, Song
    Li, Junhui
    Lu, Junzhe
    Tyagi, Rajnesh
    Liao, Zhenhua
    Feng, Pingfa
    Liu, Weiqiang
    JOURNAL OF MATERIALS ENGINEERING AND PERFORMANCE, 2017, 26 (05) : 2271 - 2284
  • [2] High cycle fatigue behavior of thermally oxidized Ti6Al4V alloy
    Cingi, Mehmet
    Meydanoglu, Onur
    Guleryuz, Hasan
    Baydogan, Murat
    Cimenoglu, Huseyin
    Kayali, E. Sabri
    PRICM 6: SIXTH PACIFIC RIM INTERNATIONAL CONFERENCE ON ADVANCED MATERIALS AND PROCESSING, PTS 1-3, 2007, 561-565 : 2179 - 2182
  • [3] Fatigue and Corrosion Fatigue of Thermally Oxidized Ti6Al4V Alloy
    dos Santos, Silvando Vieira
    Lima, Gustavo Doria
    Santos, Renan Celestino Silva
    Nascimento, Brenno Lima
    Griza, Sandro
    JOURNAL OF MATERIALS ENGINEERING AND PERFORMANCE, 2024,
  • [4] Fatigue behavior and life predictions of thermally oxidized Ti6Al4V alloy according to oxidation parameters
    Lima, Gustavo Doria
    Nascimento, Brenno Lima
    Alves Junior, Isau de Souza
    Trindade, Matheus Porto
    Griza, Sandro
    ENGINEERING FAILURE ANALYSIS, 2021, 130
  • [5] The Tribocorrosion and Corrosion Properties of Thermally Oxidized Ti6Al4V Alloy in 0.9 wt.% NaCl Physiological Saline
    Cao, Lei
    Wan, Yong
    Yang, Shuyan
    Pu, Jibin
    COATINGS, 2018, 8 (08):
  • [6] Thermal oxidation of Ti6Al4V alloy and its biotribological properties under serum lubrication
    Luo, Yong
    Chen, Wenwen
    Tian, Maocai
    Teng, Shuhua
    TRIBOLOGY INTERNATIONAL, 2015, 89 : 67 - 71
  • [7] Surface characterization and mechanical property evaluation of thermally oxidized Ti-6Al-4V
    Biswas, Amit
    Majumdar, Jyotsna Dutta
    MATERIALS CHARACTERIZATION, 2009, 60 (06) : 513 - 518
  • [8] Plasma nitriding behavior of Ti6Al4V orthopedic alloy
    Yildiz, F.
    Yetim, A. F.
    Alsaran, A.
    Celik, A.
    SURFACE & COATINGS TECHNOLOGY, 2008, 202 (11) : 2471 - 2476
  • [9] Nitriding Behavior of Ti6Al4V Alloy in Gas Atmosphere
    Siyahjani, Farid
    Atar, Erdem
    TMS 2014 SUPPLEMENTAL PROCEEDINGS, 2014, : 197 - 203
  • [10] Wear studies on ZrO2-filled PEEK as coating bearing materials for artificial cervical discs of Ti6Al4V
    Song, Jian
    Liu, Yuhong
    Liao, Zhenhua
    Wang, Song
    Tyagi, Rajnesh
    Liu, Weiqiang
    MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2016, 69 : 985 - 994