Reformatted method for two-dimensional detector arrays measurement data in proton pencil beam scanning

被引:0
作者
Meng-Ya Guo
Xiu-Fang Li
Jie Wang
Qi Liu
Xiu-Zhen Deng
Man-Zhou Zhang
Li-Ren Shen
Yue-Hu Pu
Zhi-Ling Chen
机构
[1] Chinese Academy of Sciences,Shanghai Institute of Applied Physics
[2] University of Chinese Academy of Sciences,undefined
[3] Shanghai APACTRON Particle Equipment Co. Ltd,undefined
[4] Shanghai Advanced Research Institute,undefined
[5] Chinese Academy of Sciences,undefined
来源
Nuclear Science and Techniques | 2021年 / 32卷
关键词
2D ion chamber array detectors; Laplacian pyramid image blending; High-resolution reformatted methods; Pencil beam scanning; Proton therapy;
D O I
暂无
中图分类号
学科分类号
摘要
The spatial resolution of a commercial two-dimensional (2D) ionization chamber (IC) array is limited by the size of the individual detector and the center-to-center distance between sensors. For dose distributions with areas of steep dose gradients, inter-detector dose values are derived by the interpolation of nearby detector readings in the conventional mathematical interpolation of 2D IC array measurements. This may introduce significant errors, particularly in proton spot scanning radiotherapy. In this study, by combining logfile-based reconstructed dose values and detector measurements with the Laplacian pyramid image blending method, a novel method is proposed to obtain a reformatted dose distribution that provides an improved estimation of the delivered dose distribution with high spatial resolution. Meanwhile, the similarity between the measured original data and the downsampled logfile-based reconstructed dose is regarded as the confidence of the reformatted dose distribution. Furthermore, we quantify the performance benefits of this new approach by directly comparing the reformatted dose distributions with 2D IC array detector mathematically interpolated measurements and original low-resolution measurements. The result shows that this new method is better than the mathematical interpolation and achieves gamma pass rates similar to those of the original low-resolution measurements. The reformatted dose distributions generally yield a confidence exceeding 95%.
引用
收藏
相关论文
共 71 条
  • [1] Pidikiti R(2018)Commissioning of the world's first compact pencil-beam scanning proton therapy system J. Appl. Clin. Med. Phys. 19 94-105
  • [2] Patel BC(2016)Proton beam therapy in Japan: current and future status Jap. J. Clinical Oncol. 46 885-892
  • [3] Maynard MR(2015)Current situation and perspectives of proton therapy Cancer Radiotherapie. 19 211-219
  • [4] Sakurai H(1995)The 200-MeV proton therapy project at the Paul Scherrer Institute: conceptual design and practical realization Med. Phys. 22 37-53
  • [5] Ishikawa H(2004)Treatment planning and verification of proton therapy using spot scanning initial experience Med. Phys. 31 3150-3157
  • [6] Okumura T(2004)Development of an inorganic scintillating mixture for proton beam verification dosimetry Phys. Med. Biol. 49 4637-4655
  • [7] Doyen J(2013)Preliminary investigations on the determination of three-dimensional dose distributions using scintillator blocks and optical tomography Med. Phys. 40 082104-123
  • [8] Bondiau PY(2017)Patient-Specific QA of Spot-Scanning Proton Beams using Radiochromic Film, International journal of medical physics Clin. Eng. Radiat. Oncol. 6 111-5837
  • [9] Benezery K(2010)Verification of patient-specific dose distributions in proton therapy using a commercial two-dimensional ion chamber array Med. Phys. 37 5831-3894
  • [10] Pedroni E(2008)Use of a two-dimensional ionization chamber array for proton therapy beam quality assurance Med. Phys. 35 3889-4601