Computer model for the optimization of AV and VV delay in cardiac resynchronization therapy

被引:0
作者
Matthias Reumann
Dima Farina
Raz Miri
Stephan Lurz
Brigitte Osswald
Olaf Dössel
机构
[1] IBM TJ Watson Research Center,Computational Biology Center
[2] Universität Karlsruhe (TH),Institute of Biomedical Engineering
[3] University of Heidelberg,Department of Cardiac Surgery
来源
Medical & Biological Engineering & Computing | 2007年 / 45卷
关键词
Heart failure; Cardiac resynchronization therapy; Pacing; Optimization; Computer models;
D O I
暂无
中图分类号
学科分类号
摘要
An optimal electrode position, atrio-ventricular (AV) and interventricular (VV) delay in cardiac resynchronization therapy (CRT) improves its success. An optimization strategy does not yet exist. A computer model of the Visible Man and a patient heart was used to simulate an atrio-ventricular and a left bundle branch block with 0%, 20% and 40% reduction in interventricular conduction velocity, respectively. The minimum error between physiological excitation and pathology/therapy was automatically computed for 12 different electrode positions. AV and VV delay timing was adjusted accordingly. The results show the importance of individually adjusting the electrode position as well as the timing delays to the patient’s anatomy and pathology, which is in accordance with current clinical studies. The presented methods and strategy offer the opportunity to carry out non-invasive, automatic optimization of CRT preoperatively. The model is subject to validation in future clinical studies.
引用
收藏
页码:845 / 854
页数:9
相关论文
共 82 条
  • [1] Abraham WT(2002)Cardiac resynchronization in chronic heart failure N Engl J Med 346 1845-1853
  • [2] Fisher WG(2005)Computer-aided evaluation and optimisation of biventricular pacing for patients with congestive heart failure Biomedizinische Technik 50 701-702
  • [3] Smith AL(1999)Effect of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure. The Pacing Therapies for Congestive Heart Failure Study Group. The Guidant Congestive Heart Failure Research Group Circulation 99 2993-3001
  • [4] Albrecht K(2004)Echocardiographic parameters of ventricular dyssynchrony validation in patients with heart failure using sequential biventricular pacing J Am Coll Cardiol 44 2157-2165
  • [5] Reumann M(2004)Comparison of medical therapy, pacing and defibrillation in heart failure (COMPANION) investigators. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure N Engl J Med 250 2140-2150
  • [6] Seemann G(2001)Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay N Engl J Med 344 873-880
  • [7] Auricchio A(2005)The effect of cardiac resynchronization on morbidity and mortality in heart failure N Engl J Med 352 1539-1549
  • [8] Stellbrink C(2003)Cardiac resynchronization therapy for the treatment of heart failure in patients with intraventricular conduction delay and malignant ventricular tachyarrhythmias J Am Coll Cardiol 42 1454-1459
  • [9] Block M(1999)Improved left ventricular mechanics from acute VDD pacing in patients with dilated cardiomyopathy and ventricular conduction delay Circulation 99 1567-1573
  • [10] Bordachar P(2007)Coupling of a 3D finite element model of cardiac ventricular mechanics to lumped systems models of the systemic and pulmonic circulation Ann Biomed Eng 35 1-18