Connector sensors for permittivity-based thrombus monitoring in extracorporeal life support

被引:0
作者
Martin W. Sifuna
Madoka Koishi
Takuo Uemura
Hanako Tatekawa
Hideaki Haneishi
Achyut Sapkota
Masahiro Takei
机构
[1] Chiba University,Division of Fundamental Engineering, Department of Mechanical Engineering, Graduate School of Science and Engineering
[2] Chiba University,Department of Medical Engineering, Graduate School of Science and Engineering
[3] Chiba University,Center for Frontier Medical Engineering
[4] National Institute of Technology,Department of Information and Computer Engineering
[5] Kisarazu College,undefined
来源
Journal of Artificial Organs | 2021年 / 24卷
关键词
Connector sensor; Connector–tube interface; Extracorporeal circulation; Thrombus monitoring;
D O I
暂无
中图分类号
学科分类号
摘要
Extracorporeal circulation is vital in cardiovascular surgery, but thrombus formation at connector interface is a major threat. Optical coherence tomography (OCT) is presently used to monitor thrombogenesis at connectors, but it is expensive to install and complex to use. This study fabricated and evaluated a connector sensor for real-time permittivity-based thrombus monitoring at tube–connector interface. Computational simulations were initially done to pre-evaluate the applicability of connector sensor. The sensor was fabricated by incorporating two stainless steel electrodes on acrylic tube for measuring permittivity changes at the tube–connector interface. OCT images were also taken from the interface at intervals for comparisons. Results show that the sensor was able to detect thrombus formation at the interface in form of sudden rise in permittivity after time t = 9 min. The permittivity changes were confirmed by OCT images which showed thrombus formation after time t = 14 min implying that permittivity changes were due to regional aggregation of red blood cells. The connector sensor is therefore envisioned as an affordable alternative to OCT for real-time permittivity-based monitoring of thrombogenesis at tube–connector interface.
引用
收藏
页码:15 / 21
页数:6
相关论文
共 91 条
[11]  
Otto CM(2015)Application of electrical resistance tomography for thrombus visualization in blood Flow Meas Instrum 46 334-340
[12]  
Guyton RA(2006)Assessment of coronary arterial thrombus by optical coherence tomography Am J Cardiol 97 1713-1717
[13]  
Gara PTO(2010)Dielectric coagulometry: a new approach to estimate venous thrombosis risk Anal Chem 82 9769-9774
[14]  
Sorajja P(2015)Principles of dielectric blood coagulometry as a comprehensive coagulation test Anal Chem 87 10072-10079
[15]  
Shiba N(2015)Analysis of the influence of volume and red blood cell concentration of a thrombus on the permittivity of blood J Biorheol 29 15-18
[16]  
Shimokawa H(2015)Relative permittivity measurement during the thrombus formation process using the dielectric relaxation method for various hematocrit values J Artif Organs 18 346-353
[17]  
Wasywich CA(2018)Quantitative evaluation of electrical parameters influenced by blood flow rate with multiple-frequency measurement based on modified Hanai mixture formula Sensors Actuators B Chem 268 7-14
[18]  
Gamble GD(2018)Red blood cells aggregability measurement of coagulating blood in extracorporeal circulation system with multiple-frequency electrical impedance spectroscopy Biosens Bioelectron 112 79-85
[19]  
Whalley GA(2016)Cell electrofusion in centrifuged erythrocyte pellets assessed by dielectric spectroscopy J Membr Biol 249 31-39
[20]  
Doughty RN(2018)Linear relationship between cytoplasm resistance and hemoglobin in red blood cell hemolysis by electrical impedance spectroscopy & eight-parameter equivalent circuit Biosens Bioelectron 119 103-109