A new balloon catheter for Selective Cerebral Perfusion

被引:0
|
作者
Higami T. [1 ]
Ugawa J. [2 ]
Ishihara K. [2 ]
Watanabe F. [2 ]
Sasagawa M. [2 ]
机构
[1] Division of Cardiovascular Surgery, Hyogo Brain and Heart Center, Himeji 670-0981
[2] Fuji Systems Co., Ltd., Tokyo
关键词
Aortic arch operation; Cerebral protection; Rib-surface balloon; Selective cerebral perfusion; SP Stud catheter;
D O I
10.1007/BF02479992
中图分类号
学科分类号
摘要
Selective cerebral perfusion (SCP) is a better method than hypothermic circulatory arrest for brain protection in aortic arch surgery because of fewer time limitations; however, it is more troublesome during surgery. We developed a new catheter for SCP, the SP Stud catheter, with a rib-surface balloon to reduce slipping to eliminate the need for snaring or clamping. To evaluate the slipping resistance of the new catheter, we compared it with two different balloon catheters: a Foley balloon catheter with a smooth-surface balloon as long as the SP Stud catheter, and a Retro-TH catheter with a smooth-surface balloon of discoid shape half as long as the SP Stud catheter. The "draw-strength," which refers to the slipping resistance, was measured in 20 autopsy carotid arteries in the three groups. The average draw-strengths of SP Stud catheter, Foley balloon catheter, and Retro-TH catheter were 92.3 ± 4.0, 20.8 ± 1.5, and 17.9 ± 0.8g, respectively, in sclerotic carotid arteries. The SP Stud catheter demonstrated approximately four times more resistance to slipping than the Foley balloon or Retro-TH catheters. The SP Stud catheter demonstrated a higher slipping resistance than conventional catheters, which may allow selective cerebral perfusion without snaring or clamping the carotid arteries, minimize the chance of cerebral embolism, and clear the operative field. © The Japanese Society for Artificial Organs 1998.
引用
收藏
页码:91 / 93
页数:2
相关论文
共 50 条
  • [1] PERFUSION BALLOON CATHETER
    STACK, RS
    QUIGLEY, PJ
    COLLINS, G
    PHILLIPS, HR
    AMERICAN JOURNAL OF CARDIOLOGY, 1988, 61 (14): : G77 - G80
  • [2] SELECTIVE INTERNAL ORGAN HYPERTHERMIA UTILIZING BALLOON CATHETER PERFUSION
    CHO, KJ
    ADAMS, DF
    INVESTIGATIVE RADIOLOGY, 1980, 15 (05) : 390 - 390
  • [3] A new double lumen balloon catheter for retrograde cerebral perfusion via jugular vein cannulation
    Doi, K
    Satoh, S
    Hirai, J
    Niu, S
    Kanda, K
    Okano, T
    Masuda, S
    Oka, T
    ASAIO JOURNAL, 1997, 43 (05) : M731 - M735
  • [4] Malposition of selective cerebral perfusion catheter is not a rare event
    Orihashi, K
    Sueda, T
    Okada, K
    Imai, K
    EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2005, 27 (04) : 644 - 648
  • [5] A new cannula for antegrade selective cerebral perfusion
    Venturini, Andrea
    Gallingani, Alan
    Asta, Angiolino
    Cannarella, Antonio Maria
    Zanchettin, Chiara
    Zoffoli, Giampaolo
    Mangino, Domenico
    JTCVS TECHNIQUES, 2020, 4 : 10 - 13
  • [6] THE PERFUSION BALLOON CATHETER - A NEW METHOD FOR SAFE PROLONGED CORONARY DILATATION
    COLLINS, GJ
    RAMIREZ, NM
    HINOHARA, T
    KLOPOVIC, Z
    PHILLIPS, HR
    STACK, RS
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 9 (02) : A106 - A106
  • [7] SIGNIFICANCE OF BALLOON CATHETER SELECTIVE AND SUPERSELECTIVE ANGIOGRAPHY IN THE DIAGNOSIS OF CEREBRAL VASCULAR MALFORMATIONS
    CZIRJAK, S
    PASZTOR, E
    LAZAR, L
    DEAK, G
    LANYI, F
    ACTA NEUROCHIRURGICA, 1981, 58 (1-2) : 85 - 98
  • [8] Vascular Dynamics of Cerebral Gliomas Investigated with Selective Catheter Angiography, Perfusion CT and MRI
    Wiest, Roland
    Schaer, Ralph
    Reinert, Michael H.
    von Bredow, Ferdinand H.
    El Koussy, Marwan M.
    Remonda, Luca
    Schroth, Gerhard
    Ozdoba, Christoph
    Slotboom, Johannes
    CLINICAL NEURORADIOLOGY, 2008, 18 (02) : 98 - 106
  • [9] FRACTURE OF A PERFUSION BALLOON CATHETER DURING CORONARY ANGIOPLASTY
    KUAN, PL
    KO, YL
    TSAI, CH
    LIN, JL
    CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1994, 32 (01): : 33 - 35
  • [10] BALLOON CATHETER AS A MODEL OF CEREBRAL EMBOLI IN HUMANS
    GACS, G
    MEREI, FT
    BODOSI, M
    STROKE, 1982, 13 (01) : 39 - 42