Monte Carlo study of conversion factors for ionization chamber dosimetry in solid slab phantoms for MV photon beams

被引:0
|
作者
Dong-wook Park
Jai-ki Lee
机构
[1] Hanyang University,Department of Nuclear Engineering
[2] Ilsan Paik Hospital,Department of Radiation Oncology
来源
Journal of the Korean Physical Society | 2016年 / 69卷
关键词
Monte Carlo; Solid phantom; RW3; Ion chamber; FC65-G;
D O I
暂无
中图分类号
学科分类号
摘要
For high energy photon beams, solid phantom to water dose conversion factors were calculated by using a Monte Carlo method, and the result were compared with measurements and published data. Based on the absorbed dose to water dosimetry protocol, the conversion factor was theoretically divided into stopping powers ratios, perturbation factors and ratios of absorbed dose to water and that to solid phantom. Data for a Farmer-type chamber and a solid phantom based on polystyrene which is one of the most common material were applied to calculate the conversion factors for 6 MV and 15 MV photon beams. All measurements were conducted after 10 Gy pre-irradiation and thermal equilibrium had been established with solid slabs in a treatment room. The calculated and the measured conversion factors were in good agreement and could be used to confirm the feasibility of the solid phantom as a substitute for water for high energy photon beam.
引用
收藏
页码:647 / 652
页数:5
相关论文
共 50 条
  • [21] Monte Carlo optimization and experimental validation of a prototype ionization chamber for accurate magnetic resonance image guided radiation therapy (MRgRT) daily output constancy measurements in solid phantoms
    Muir, Bryan R.
    Nusrat, Humza
    Sarfehnia, Arman
    Renaud, James
    MEDICAL PHYSICS, 2022, 49 (08) : 5483 - 5490
  • [22] A new reference-type ionization chamber with direction-independent response for use in small-field photon-beam dosimetry - An experimental and Monte Carlo study
    Delfs, Bjoern
    Kapsch, Ralf-Peter
    Chofor, Ndimofor
    Looe, Hui Khee
    Harder, Dietrich
    Poppe, Bjoern
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2019, 29 (01): : 39 - 48
  • [23] Monte Carlo calculation of photo-neutron dose produced by circular cones at 18 MV photon beams
    Hosseinzadeh, Elham
    Banaee, Nooshin
    Nedaie, Hassan Ali
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2018, 23 (01) : 39 - 46
  • [24] A Simple Source Model for 6 MV Flattening Filter Free Photon Beams Monte Carlo Dose Calculations
    Matthew T. Ahad Ollah Ezzati
    Negin Studenski
    Physics of Particles and Nuclei Letters, 2021, 18 : 791 - 798
  • [25] A Simple Source Model for 6 MV Flattening Filter Free Photon Beams Monte Carlo Dose Calculations
    Ezzati, Ahad Ollah
    Studenski, Matthew T.
    Jamshidi, Negin
    PHYSICS OF PARTICLES AND NUCLEI LETTERS, 2021, 18 (07) : 791 - 798
  • [26] Dose measurements compared with Monte Carlo simulations of narrow 6 MV multileaf collimator shaped photon beams
    De Vlamynck, K
    Palmans, H
    Verhaegen, F
    De Wagter, C
    De Neve, W
    Thierens, H
    MEDICAL PHYSICS, 1999, 26 (09) : 1874 - 1882
  • [27] Evaluation of an extrapolation chamber for dosimetry in computed tomography beams using Monte Carlo code (MCNP5)
    Castro, Maysa C.
    Silva, Natalia F.
    Santos, Lucas R.
    Cintra, Felipe B.
    Caldas, Linda V. E.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 167
  • [28] Determining superficial dosimetry for the internal canthus from the Monte Carlo simulation of kV photon and MeV electron beams
    B. E. Currie
    Australasian Physics & Engineering Sciences in Medicine, 2009, 32 : 68 - 80
  • [29] Determining superficial dosimetry for the internal canthus from the Monte Carlo simulation of kV photon and MeV electron beams
    Currie, B. E.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2009, 32 (02) : 68 - 80
  • [30] A feasibility study to calculate unshielded fetal doses to pregnant patients in 6-MV photon treatments using Monte Carlo methods and anatomically realistic phantoms
    Bednarz, Bryan
    Xu, X. George
    MEDICAL PHYSICS, 2008, 35 (07) : 3054 - 3061