The reduction in Monte Carlo calculated organ doses from CT with tube current modulation using WILLIAM, a voxel model of seven year-old anatomy

被引:0
作者
Martin Caon
机构
[1] Flinders University,School of Health Sciences
[2] Flinders University,Medical Devices Research Institute
来源
Australasian Physical & Engineering Sciences in Medicine | 2014年 / 37卷
关键词
Tube current modulation; Voxel model; Organ dose; Monte Carlo; Computed tomography; AEC;
D O I
暂无
中图分类号
学科分类号
摘要
The construction of a voxel model of seven year-old male anatomy from a low resolution whole body CT scan acquired with a PET-CT scanner is described. The model is referred to as WILLIAM. The model was used to compare the Monte Carlo calculated relative organ doses with and without tube current modulation from simulated CAP and chest CT examinations. Tube current modulation was simulated in both the angular (x–y plane) and axial (along the z axis) senses. For organs that are located wholly within the scanned anatomy, significant relative dose reductions of approximately 43 to 62 % were calculated when tube current modulation was simulated compared to Monte Carlo calculations that did not include tube current modulation. Organs located outside of the directly irradiated anatomy, experienced a decrease in their relative dose of between 33 and 38 % when tube current modulation was simulated.
引用
收藏
页码:743 / 752
页数:9
相关论文
共 87 条
[1]  
Zankl M(1988)The construction of computer tomographic phantoms and their application in radiology and radiation protection Radiat Environ Biophys 27 153-164
[2]  
Veit R(1999)An EGS4-ready tomographic computational model of a fourteen year-old female torso for calculating organ doses from CT examinations Phys Med Biol 44 2213-2225
[3]  
Williams G(2002)Creation of two tomographic voxel models of paediatric patients in the first year of life Phys Med Biol 47 3143-3164
[4]  
Schneider K(2005)The UF series of tomographic computational phantoms of pediatric patients Med Phys 32 3537-3548
[5]  
Fendel H(2010)The UF family of reference hybrid phantoms for computational radiation dosimetry Phys Med Biol 55 339-363
[6]  
Petoussi N(2012)Dose conversion coefficients for paediatric CT examinations with automatic tube current modulation Phys Med Biol 57 6309-6326
[7]  
Drexler G(2006)Development of the two Korean adult tomographic computational phantoms for organ dosimetry Med Phys 33 380-390
[8]  
Caon M(2009)Development of 7-year-old Korean child model for computational dosimetry ETRI J 31 237-239
[9]  
Bibbo G(2010)The virtual family - development of surface-based anatomical models of two adults and two children for dosimetric simulations Phys Med Biol 55 N23-N38
[10]  
Pattison J(2000)Monte Carlo calculated effective dose to teenage girls from CT examinations Radiat Prot Dosimetry 90 445-448