共 50 条
Late type III endoleak from fabric tears of a zenith stent graft: report of a case
被引:0
|作者:
Hiroshi Banno
Hirofumi Morimae
Tsutomu Ihara
Masayoshi Kobayashi
Kiyohito Yamamoto
Kimihiro Komori
机构:
[1] Nagoya University Graduate School of Medicine,Division of Vascular Surgery, Department of Surgery
来源:
Surgery Today
|
2012年
/
42卷
关键词:
Endoleak;
Fabric tears;
Stent graft;
EVAR;
D O I:
暂无
中图分类号:
学科分类号:
摘要:
We report a case of late type III B endoleak from a fabric tear in the main body of a Zenith bifurcated stent-graft approximately 19 months after implantation. A follow-up computed tomography (CT) scan showed gradual growth in the size of the aneurysm, but no apparent endoleak was recognized. However, the CT scan had demonstrated a mosaic pattern in the aneurysm sac, which could indicate an endoleak from a fabric tear. The defect was repaired by implantation of body extensions, but the patient died 5 h after the procedure.
引用
收藏
页码:1206 / 1209
页数:3