A Technical Assessment of Pulse Wave Velocity Algorithms Applied to Non-invasive Arterial Waveforms

被引:0
作者
N. R. Gaddum
J. Alastruey
P. Beerbaum
P. Chowienczyk
T. Schaeffter
机构
[1] King’s College London,Division of Imaging Sciences and Biomedical Engineering, St. Thomas’ Hospital
[2] Hannover Medical School,Department of Pediatric Cardiology & Pediatric Intensive Care Medicine
[3] King’s College London British Heart Foundation Centre,St Thomas’ Hospital
来源
Annals of Biomedical Engineering | 2013年 / 41卷
关键词
Pulse wave velocity; Foot to foot; Least squares; Cross correlation; Doppler ultrasound; Tonometry; One-dimensional modelling;
D O I
暂无
中图分类号
学科分类号
摘要
Non-invasive assessment of arterial stiffness through pulse wave velocity (PWV) analysis is becoming common clinical practice. However, the effects of measurement noise, temporal resolution and similarity of the two waveforms used for PWV calculation upon accuracy and variability are unknown. We studied these effects upon PWV estimates given by foot-to-foot, least squared difference, and cross-correlation algorithms. We assessed accuracy using numerically generated blood pressure and flow waveforms for which the theoretical PWV was known to compare with the algorithm estimates. We assessed variability using clinical measurements in 28 human subjects. Wave shape similarity was quantified using a cross correlation-coefficient (CCCoefficient), which decreases with increasing distance between waveform measurements sites. Based on our results, we propose the following criteria to identify the most accurate and least variable algorithm given the noise, resolution and CCCoefficient of the measured waveforms. (1) Use foot-to-foot when the noise-to-signal ratio ≤10%, and/or temporal resolution ≥100 Hz. Otherwise (2) use a least squares differencing method applied to the systolic upstroke.
引用
收藏
页码:2617 / 2629
页数:12
相关论文
共 50 条
  • [1] A Technical Assessment of Pulse Wave Velocity Algorithms Applied to Non-invasive Arterial Waveforms
    Gaddum, N. R.
    Alastruey, J.
    Beerbaum, P.
    Chowienczyk, P.
    Schaeffter, T.
    ANNALS OF BIOMEDICAL ENGINEERING, 2013, 41 (12) : 2617 - 2629
  • [2] Robust and practical non-invasive estimation of local arterial wave speed and mean blood velocity waveforms
    Kowalski, Remi
    Beare, Richard
    Willemet, Marie
    Alastruey, Jordi
    Smolich, Joseph J.
    Cheung, Michael M. H.
    Mynard, Jonathan P.
    PHYSIOLOGICAL MEASUREMENT, 2017, 38 (11) : 2081 - 2099
  • [3] Limitations and pitfalls of non-invasive measurement of arterial pressure wave reflections and pulse wave velocity
    Segers P.
    Kips J.
    Trachet B.
    Swillens A.
    Vermeersch S.
    Mahieu D.
    Rietzschel E.
    De Buyzere M.
    Van Bortel L.
    Artery Research, 2009, 3 (2) : 79 - 88
  • [4] Non-invasive assessment of pulse wave velocity in mice by means of ultrasound images
    Di Lascio, Nicole
    Stea, Francesco
    Kusmic, Claudia
    Sicari, Rosa
    Faita, Francesco
    ATHEROSCLEROSIS, 2014, 237 (01) : 31 - 37
  • [5] Validation of a new non-invasive portable tonometer for determining arterial pressure wave and pulse wave velocity: the PulsePen device
    Salvi, P
    Lio, G
    Labat, C
    Ricci, E
    Pannier, B
    Benetos, A
    JOURNAL OF HYPERTENSION, 2004, 22 (12) : 2285 - 2293
  • [6] Non-invasive assessment of ventricular-arterial coupling: correlation between myocardial work and the pulse wave velocity parameters
    Parlavecchio, Antonio
    Caminiti, Rodolfo
    Vetta, Giampaolo
    Magnocavallo, Michele
    Della Rocca, Domenico Giovanni
    Foti, Rosario
    Ajello, Manuela
    Crea, Pasquale
    Dattilo, Giuseppe
    Siviglia, Massimo
    Postorino, Stefano
    Pangallo, Antonio
    Benedetto, Francesco Antonio
    Zito, Concetta
    Carerj, Scipione
    COR ET VASA, 2023, 65 (01) : 51 - 56
  • [7] Non-Invasive Assessment of Arterial Stiffness: Pulse Wave Velocity, Pulse Wave Analysis and Carotid Cross-Sectional Distensibility: Comparison between Methods
    Salvi, Paolo
    Valbusa, Filippo
    Kearney-Schwartz, Anna
    Labat, Carlos
    Grillo, Andrea
    Parati, Gianfranco
    Benetos, Athanase
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (08)
  • [8] Non-invasive measures of pulse wave velocity correlate with coronary arterial plaque load in humans
    McLeod, AL
    Uren, NG
    Wilkinson, IB
    Webb, DJ
    Maxwell, SRJ
    Northridge, DB
    Newby, DE
    JOURNAL OF HYPERTENSION, 2004, 22 (02) : 363 - 368
  • [9] ASSESSMENT OF PULSE WAVE VELOCITY: COMPARISON OF A NON-INVASIVE OPTICAL METHOD TO A STANDARD COMMERCIAL DEVICE (THE COMPLIOR SP)
    Horsley, Kristin
    Liu, Manlu
    Ditto, Blaine
    PSYCHOPHYSIOLOGY, 2018, 55 : S58 - S58
  • [10] Repeatability of non-invasive measurement of intracerebral pulse wave velocity using transcranial Doppler
    Gladdish, S
    Manawadu, D
    Banya, W
    Cameron, J
    Bulpitt, CJ
    Rajkumar, C
    CLINICAL SCIENCE, 2005, 108 (05) : 433 - 439