A Method for Methylmercury and Inorganic Mercury in Biological Samples Using High Performance Liquid Chromatography- Inductively Coupled Plasma Mass Spectrometry

被引:0
|
作者
Tomohiro Narukawa
Takahiro Iwai
Koichi Chiba
Joerg Feldmann
机构
[1] National Institute of Advanced Industrial Science and Technology (AIST),National Metrology Institute of Japan (NMIJ)
[2] School of Science and Technology,Department of Environmental and Applied Chemistry
[3] Kwansei Gakuin University,undefined
[4] Environmental Analytical Chemistry TESLA-Trace Element Speciation Laboratory,undefined
[5] University of Aberdeen,undefined
来源
Analytical Sciences | 2018年 / 34卷
关键词
Keywords Mercury speciation; HPLC-ICP-MS; methylmercury determination; biological samples; risk assessment; alkaline extraction;
D O I
暂无
中图分类号
学科分类号
摘要
A new determination method was developed for the measurement of methylmercury (Me-Hg) and inorganic mercury (i-Hg) in biological samples using high-performance liquid chromatography-inductively coupled plasma-mass spectrometry (HPLC-ICP-MS) following alkaline extraction. Mercury species in biological samples were extracted with 10% (w/w) tetramethylammonium hydroxide (TMAH) solution at 80°C for 2 h. Methylmercury was completely separated from i-Hg by adamantyl type and octadecylsilyl type columns within 6 and 4 min using isocratic elution, respectively. The detection limits (3(3σ)) of adamantyl and octadecylsilyl columns using the proposed system were 0.08 and 0.13 ng g−1 (as Hg), respectively. Inorganic Hg completely separates from Me-Hg without tailing. The proposed determination methods were applied to several biological certified reference materials (CRMs). The measurement results of Me-Hg obtained by the present method were in good agreement within the expanded uncertainties (k = 2) with the certified values. The analytical precision (n = 3) of Me-Hg was less than 2%, and the recoveries of Me-Hg and i-Hg were 101 ± 1 and 103 ± 3%, respectively. In addition, this method enables the determination of Me-Hg and i-Hg for 20 samples in 1 h.
引用
收藏
页码:1329 / 1334
页数:5
相关论文
共 50 条
  • [1] A Method for Methylmercury and Inorganic Mercury in Biological Samples Using High Performance Liquid Chromatography Inductively Coupled Plasma Mass Spectrometry
    Narukawa, Tomohiro
    Iwai, Takahiro
    Chiba, Koichi
    Feldmann, Joerg
    ANALYTICAL SCIENCES, 2018, 34 (11) : 1329 - 1334
  • [2] Determination of inorganic and methyl mercury in biological samples by high performance liquid chromatography coupled with inductively coupled plasma-mass spectrometry
    Wang, M
    Feng, WY
    Zhang, F
    Wang, B
    Shi, JW
    Li, B
    Chai, ZF
    Zhao, YL
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2005, 33 (12) : 1671 - 1675
  • [3] A simple method for methylmercury, inorganic mercury and ethylmercury determination in plasma samples by high performance liquid chromatography-cold-vapor-inductively coupled plasma mass spectrometry
    de Souza, Samuel S.
    Campiglia, Andres Dobal
    Barbosa, Fernando, Jr.
    ANALYTICA CHIMICA ACTA, 2013, 761 : 11 - 17
  • [4] The speciation of inorganic and methylmercury in human hair by high-performance liquid chromatography coupled with inductively coupled plasma mass spectrometry
    Morton, J
    Carolan, VA
    Gardiner, PHE
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (04) : 377 - 381
  • [5] Problems encountered during the development of a method for the speciation of mercury and methylmercury by high-performance liquid chromatography coupled to inductively coupled plasma mass spectrometry
    Harrington, CF
    Catterick, T
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (09) : 1053 - 1056
  • [6] Development of a liquid chromatography-inductively coupled plasma mass spectrometry method for the simultaneous determination of methylmercury and inorganic mercury in human blood
    Sogame, Yoshihisa
    Tsukagoshi, Akihiko
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2020, 1136
  • [7] Methylmercury and inorganic mercury determination in blood by using liquid chromatography with inductively coupled plasma mass spectrometry and a fast sample preparation procedure
    Rodrigues, Jairo L.
    de Souza, Samuel S.
    de Oliveira Souza, Vanessa C.
    Barbosa, Fernando, Jr.
    TALANTA, 2010, 80 (03) : 1158 - 1163
  • [8] Methylmercury Detection in Maternal Blood Samples by Liquid Chromatography with Inductively Coupled Plasma Mass Spectrometry
    Abedinlah, Amirah
    Elias, Saliza Mohd
    Praveena, Sarva Mangala
    Abu Bakar, Suhaili
    Rejali, Zulida
    Jalaludin, Juliana
    PERTANIKA JOURNAL OF SCIENCE AND TECHNOLOGY, 2021, 29 (03): : 2027 - 2044
  • [9] Determ ination of methylmercury in the tuna tissue by High performance liquid chromatography with Inductively Coupled Plasma Mass Spectrometry
    LV Lei1 Jiang Zhigang 2 Li Qun1 Zhang Shuai2 (1.The Department of the Application Chemistey Qingdao University 2.Qingdao Entry-Exit Inspection And Quarantine Burea)
    科技信息, 2009, (06) : 472 - 473
  • [10] Antimony speciation in environmental samples by using high-performance liquid chromatography coupled to inductively coupled plasma mass spectrometry
    Zheng, JA
    Ohata, M
    Furuta, N
    ANALYTICAL SCIENCES, 2000, 16 (01) : 75 - 80