Application of Zeeman Graphite Furnace Atomic Absorption Spectrometry with High-Frequency Modulation Polarization for the Direct Determination of Aluminum, Beryllium, Cadmium, Chromium, Mercury, Manganese, Nickel, Lead, and Thallium in Human Blood

被引:0
|
作者
Natalya B. Ivanenko
Nikolay D. Solovyev
Anatoly A. Ivanenko
Alexander A. Ganeev
机构
[1] Federal Medico-Biological Agency of Russia,Institute of Toxicology
[2] Russian Federation,undefined
[3] Saint Petersburg State University,undefined
[4] Russian Federation,undefined
来源
Archives of Environmental Contamination and Toxicology | 2012年 / 63卷
关键词
AuCl3; H2PtCl6; Flame Atomic Absorption Spectrometry; Graphite Furnace Atomic Absorption Spectrometry; Na2WO4;
D O I
暂无
中图分类号
学科分类号
摘要
Determination of aluminum (Al), beryllium (Be), cadmium (Cd), chromium (Cr), mercury (Hg), manganese (Mn), nickel (Ni), lead (Pb), and thallium (Tl) concentrations in human blood using high-frequency modulation polarization Zeeman graphite furnace atomic absorption spectrometry (GFAAS) was performed. No sample digestion was used in the current study. Blood samples were diluted with deionized water or 0.1 % (m/v) Triton X-100 solution for Tl. Dilution factors ranged from 1/5 per volume for Be and Tl to 1/20 per volume for Cd and Pb. For Tl, Cd, and Hg, noble metals (gold, platinum, rhodium, etc.) were applied as surface modifiers. To mitigate chloride interference, 2 % (m/v) solution of NH4NO3 was used as matrix modifier for Tl and Ni assessment. The use of Pd(NO3)2 as oxidative modifier was necessary for blood Hg and Tl measurement. Validation of the methods was performed by analyzing two-level reference material Seronorm. The precision of the designed methods as relative SD was between 4 and 12 % (middle of a dynamic range) depending on the element. For additional validation, spiked blood samples were analyzed. Limits of detection (LoDs, 3σ, n = 10) for undiluted blood samples were 2.0 μg L−1 for Al, 0.08 μg L−1 for Be, 0.10 μg L−1 for Cd, 2.2 μg L−1 for Cr, 7 μg L−1 for Hg, 0.4 μg L−1 for Mn, 2.3 μg L−1 for Ni, 3.4 μg L−1 for Pb, and 0.5 μg L−1 for Tl. The LoDs achieved allowed determination of Al, Cd, Cr, Mn, Ni, and Pb at both toxic and background levels. Be, Hg, and Tl could be reliably measured at toxic levels only. The methods developed are used for clinical diagnostics and biological monitoring of work-related exposure.
引用
收藏
页码:299 / 308
页数:9
相关论文
共 2 条
  • [1] Determination of Chromium, Copper, Iron, Manganese and Lead in Human Hair by Graphite Furnace Atomic Absorption Spectrometry
    Xiaonan Dong
    Yuzuru Nakaguchi
    Keizo Hiraki
    Analytical Sciences, 1998, 14 : 785 - 789
  • [2] Determination of chromium, copper, iron, manganese and lead in human hair by graphite furnace atomic absorption spectrometry
    Dong, XN
    Nakaguchi, Y
    Hiraki, K
    ANALYTICAL SCIENCES, 1998, 14 (04) : 785 - 789