Use of a megavoltage electronic portal imaging device to identify prosthetic materials

被引:0
作者
Vaughan Moutrie
Tanya Kairn
Anatoly Rosenfeld
Paul H. Charles
机构
[1] Princess Alexandra Hospital,Department of Radiation Oncology
[2] Premion,Centre for Medical Radiation Physics
[3] The Wesley Medical Centre,undefined
[4] University of Wollongong,undefined
来源
Australasian Physical & Engineering Sciences in Medicine | 2015年 / 38卷
关键词
EPID; Megavoltage imaging; Prosthesis; Radiation therapy; Electron density;
D O I
暂无
中图分类号
学科分类号
摘要
To achieve accurate dose calculations in radiation therapy the electron density of patient tissues must be known. This information is ordinarily gained from a computed tomography (CT) image that has been calibrated to allow relative electron density (RED) to be determined from CT number. When high density objects such as metallic prostheses are involved, direct use of the CT data can become problematic due to the artefacts introduced by high attenuation of the beam. This requires manual correction of the density values, however the properties of the implanted prosthetic are not always known. A method is introduced where the RED of such an object can be determined using the treatment beam of a linear accelerator with an electronic portal imaging device. The technique was tested using a metallic hip replacement that was placed within a container of water. Compared to the theoretical RED of 6.8 for cobalt–chromium alloy, these measurements calculated a value of 6.4 ± 0.7. This would allow the distinction of an implant as Co–Cr or steel, which have similar RED, or titanium, which is much less dense with an RED of 3.7.
引用
收藏
页码:93 / 100
页数:7
相关论文
共 101 条
[1]  
Mutic S(2003)Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: report of the AAPM Radiation Therapy Committee task group no. 66 Med Phys 30 2762-2792
[2]  
Palta J(2012)CT number definition Radiat Phys Chem 81 358-361
[3]  
Butker E(1992)An electron density calibration phantom for CT-based treatment planning computers Med Phys 19 325-327
[4]  
Das I(2006)Monte Carlo calculations of the impact of a hip prosthesis on the dose distribution Nucl Instr Meth B 251 9-18
[5]  
Huq M(2003)Calibration of CT Hounsfield units for radiotherapy treatment planning of patients with metallic hip prostheses: the use of the extended CT-scale Phys Med Biol 48 1591-1603
[6]  
Loo D(2008)Improving accuracy of electron density measurement in the presence of metallic implants using orthovoltage computed tomography Med Phys 35 1932-1941
[7]  
Salter B(2009)The adaptation of megavoltage cone beam CT for use in standard radiotherapy treatment planning Phys Med Biol 54 2067-2077
[8]  
McCollough C(2011)Validation of image registration and fusion of MV CBCT and planning CT for radiotherapy treatment planning Australas Phys Eng Sci Med 34 441-447
[9]  
Van Dyk J(2013)A snapshot of radiation therapy techniques and technology in Queensland: an aid to mapping undergraduate curriculum J Med Radiat Sci 60 25-34
[10]  
Bryant J(2003)Dosimetric considerations for patients with HIP prostheses undergoing pelvic irradiation. Report of the AAPM Radiation Therapy Committee task group 63 Med Phys 30 1162-1182