Biomechanical Comparison of Spinopelvic Fixation Constructs: Iliac Screw Versus S2-Alar-Iliac Screw

被引:73
作者
Burns C.B. [1 ]
Dua K. [1 ]
Trasolini N.A. [1 ]
Komatsu D.E. [1 ]
Barsi J.M. [1 ]
机构
[1] Department of Orthopaedic Surgery, Stony Brook University, School of Medicine, Stony Brook, 11794, NY
关键词
Biomechanical study; Iliac screw; Load-to-failure; S2-Alar-Iliac (S2AI) screw; Spinopelvic fixation;
D O I
10.1016/j.jspd.2015.07.008
中图分类号
学科分类号
摘要
Study Design Biomechanical cadaveric study. Objective To compare the biomechanical properties of the iliac and S2-Alar-Iliac (S2AI) screw in a similar spinopelvic fixation construct. Summary of Background Data Spinopelvic fixation is used in the correction of pelvic obliquity, high-grade spondylolisthesis, and long spinal fusions. With the development of pedicle screw fixation, the iliac screw has been used as an anchor point to the pelvis. The associated morbidity with this fixation has led to the development of the S2AI screw. Many studies have examined the biomechanical properties of iliac and S2AI screws; however, a direct comparison has not been performed. Methods Eight cadaveric spines were instrumented with pedicle screws bilaterally at L5 and S1. Four specimens were further instrumented with iliac screws placed with a starting point at the posterior superior iliac spine, and four specimens were instrumented with S2AI screws placed with a starting point 1 mm inferolateral to the S1 foramen. Screws were connected with 6.35 mm rods. Subfailure testing was performed by loading at 1°/second to a torque of 10 Nm in four directions: left bending, right bending, extension, and flexion. Specimens then underwent a monotonic load to failure under flexion at a rate of 1°/second. Results There were no significant differences for torsional stiffness in extension, flexion, left bending, or right bending between S2AI and iliac screw constructs. There were no significant differences in S2AI versus iliac screws for failure torque (30.9 ± 12.00 Nm vs. 22.61 ± 6.25 Nm) and yield torque (11.86 ± 0.41 Nm vs. 12.01 ± 1.70 Nm). Conclusion Iliac screws have been associated with increased dissection, wound complications, an additional construct failure point, and hardware prominence. The S2AI screw was developed as an alternative and has been associated with less morbidity. The iliac and S2AI screw demonstrate no statistical difference in stiffness and load-to-failure in a spinopelvic fixation model. Level of Evidence Level V. © 2016 Scoliosis Research Society.
引用
收藏
页码:10 / 15
页数:5
相关论文
共 22 条
[1]  
Kebaish K.M., Sacropelvic fixation: Techniques and complications, Spine (Phila Pa 1976), 35, pp. 2245-2251, (2010)
[2]  
Kuklo T.R., Bridwell K.H., Lewis S.J., Et al., Minimum 2-year analysis of sacropelvic fixation and L5-S1 fusion using S1 and iliac screws, Spine (Phila Pa 1976), 26, pp. 1976-1983, (2001)
[3]  
Chang T.L., Sponseller P.D., Kebaish K.M., Et al., Low profile pelvic fixation: Anatomic parameters for sacral alar-iliac fixation versus traditional iliac fixation, Spine (Phila Pa 1976), 34, pp. 436-440, (2009)
[4]  
Sponseller P.D., Zimmerman R.M., Ko P.S., Et al., Low profile pelvic fixation with the sacral alar iliac technique in the pediatric population improves results at two-year minimum follow-up, Spine (Phila Pa 1976), 35, pp. 1887-1892, (2010)
[5]  
Kebaish K.M., Gunne A.P.T., Mohamed A.S., Et al., A new low profile sacro-pelvic fixation using S2 alar iliac (S2AI) screws in adult deformity fusion to the sacrum: A prospective study with minimum two-year follow-up: E-poster #21, Spine, 10, (2009)
[6]  
O'Brien J.R., Yu W.D., Bhatnagar R., Et al., An anatomic study of the S2 iliac technique for lumbopelvic screw placement, Spine (Phila Pa 1976), 34, pp. E439-E442, (2009)
[7]  
O'Brien J.R., Yu W., Kaufman B.E., Et al., Biomechanical evaluation of S2 alar-iliac screws: Effect of length and quad-cortical purchase as compared with iliac fixation, Spine (Phila Pa 1976), 38, pp. E1250-E1255, (2013)
[8]  
Goel V.K., Panjabi M.M., Patwardhan A.G., Et al., Test protocols for evaluation of spinal implants, J Bone Joint Surg Am, 88, pp. 103-109, (2006)
[9]  
McCord D.H., Cunningham B.W., Shono Y., Et al., Biomechanical analysis of lumbosacral fixation, Spine (Phila Pa 1976), 17, 8, pp. S235-S243, (1992)
[10]  
Allen B.L., Ferguson R.L., The Galveston technique of pelvic fixation with L-rod instrumentation of the spine, Spine (Phila Pa 1976), 9, pp. 388-394, (1984)