A Monte Carlo evaluation of three Compton camera absorbers

被引:0
作者
C. Z. Uche
W. H. Round
M. J. Cree
机构
[1] University of Waikato,School of Engineering
来源
Australasian Physical & Engineering Sciences in Medicine | 2011年 / 34卷
关键词
Compton camera; Absorber; GEANT4; Detector parameters; Gamma rays;
D O I
暂无
中图分类号
学科分类号
摘要
We present a quantitative study on the performance of cadmium zinc telluride (CZT), thallium-doped sodium iodide (NaI(Tl)) and germanium (Ge) detectors as potential Compton camera absorbers. The GEANT4 toolkit was used to model the performance of these materials over the nuclear medicine energy range. CZT and Ge demonstrate the highest and lowest efficiencies respectively. Although the best spatial resolution was attained for Ge, its lowest ratio of single photoelectric to multiple interactions suggests that it is most prone to inter-pixel cross-talk. In contrast, CZT, which demonstrates the least positioning error due to multiple interactions, has a comparable spatial resolution with Ge. Therefore, we modelled a Compton camera system based on silicon (Si) and CZT as the scatterer and absorber respectively. The effects of the detector parameters of our proposed system on image resolution were evaluated and our results show good agreement with previous studies. Interestingly, spatial resolution which accounted for the least image degradation at 140.5 keV became the dominant degrading factor at 511 keV, indicating that the absorber parameters play some key roles at higher energies. The results of this study have validated the predictions by An et al. which state that the use of a higher energy gamma source together with reduction of the absorber segmentation to sub-millimetre could achieve the image resolution of 5 mm required in medical imaging.
引用
收藏
页码:351 / 360
页数:9
相关论文
共 98 条
  • [1] Singh M(1983)An electronically collimated gamma camera for single photon emission computed tomography. Part I: theoretical considerations and design criteria Med Phys 10 421-427
  • [2] Chelikani S(2004)Optimizing Compton camera geometries Phys Med Biol 49 1387-1408
  • [3] Gore J(1997)Doppler broadening of energy spectra in Compton cameras IEEE Nucl Sci Symp 2 1361-1365
  • [4] Zubal G(1998)Angular uncertainties due to geometry and spatial resolution in Compton cameras IEEE Nucl Sci Symp 3 1535-1540
  • [5] Ordonez CE(2002)Silicon detector for a Compton camera in nuclear medical imaging IEEE Trans Nucl Sci 49 812-816
  • [6] Bolozdynya A(2004)Silicon drift detector readout electronics for a Compton camera Nucl Instrum Methods A 523 435-440
  • [7] Chang W(2003)Development of silicon pad detectors and readout electronics for a Compton camera Nucl Instrum Methods A 501 273-279
  • [8] Ordonez CE(2008)Optimal geometrical configuration of a double-scattering Compton camera for maximum imaging resolution and sensitivity Nucl Instrum Methods A 591 80-83
  • [9] Chang W(2007)Effect of detector parameters on the image quality of Compton camera for Nucl Instrum Methods A 571 251-254
  • [10] Bolozdynya A(2004)Tc Phys Med Biol 49 617-638