Accurate scatter correction for transmission computed tomography using an uncollimated line array source

被引:0
作者
Akihiro Kojima
Masanori Matsumoto
Seiji Tomiguchi
Noboru Katsuda
Yasuyuki Yamashita
Nobutoku Motomura
机构
[1] Kumamoto University,Institute of Resource Development and Analysis
[2] Kumamoto University School of Health Sciences,Course of Radiological Sciences
[3] Kumamoto University School of Medicine,Department of Radiology
[4] Toshiba Medical Systems Corporation,undefined
来源
Annals of Nuclear Medicine | 2004年 / 18卷
关键词
attenuation correction; TEW scatter correction method; uncollimated transmission; CT source; SPECT; quantification;
D O I
暂无
中图分类号
学科分类号
摘要
We investigated scatter correction in transmission computed tomography (TCT) imaging by the combination of an uncollimated transmission source and a parallel-hole collimator. We employed the triple energy window (TEW) as the scatter correction and found that the conventional TEW method, which is accurate in emission computed tomography (ECT) imaging, needs some modification in TCT imaging based on our phantom studies. In this study a Tc-99m uncollimated line array source (area: 55 cm × 40 cm) was attached to one camera head of a dual-head gamma camera as a transmission source, and TCT data were acquired with a low-energy, general purpose (LEGP), parallel-hole collimator equipped on the other camera head. The energy spectra for 140 keV-photons transmitted through various attenuating material thicknesses were measured and analyzed for scatter fraction. The results of the energy spectra showed that the photons transmitted had an energy distribution that constructs a scatter peak within the 140 keV-photopeak energy window. In TCT imaging with a cylindrical water phantom, the conventional TEW method with triangle estimates (subtraction factor,K = 0.5) was not sufficient for accurate scatter correction (μ = 0.131 cm-1 for water), whereas the modified TEW method withK= 1.0 gave the accurate attenuation coefficient of 0.153 cm-1 for water. For the TCT imaging with the combination of the uncollimated Tc-99m line array source and parallel hole collimator, the modified TEW method withK = 1.0 gives the accurate TCT data for quantitative SPECT imaging in comparison with the conventional TEW method withK= 0.5.
引用
收藏
页码:45 / 50
页数:5
相关论文
共 123 条
[1]  
Jaszczak RJ(1984)Improved SPECT quantification using compensation for scattered photons J Nucl Med 25 893-900
[2]  
Green KL(1991)A practical method for position-dependent Compton-scatter correction in single photon emission CT IEEE Trans Med Imag 10 408-412
[3]  
Floyd CE(1993)Effect of energy resolution on scatter fraction in scintigraphic imaging: Monte Carlo study Med Phys 20 1107-1113
[4]  
Harris CC(1992)Correction of scattered photons in Tc-99m imaging by means of a photopeak dual-energy window acquisition Ann Nucl Med 6 153-158
[5]  
Coleman RE(1993)Compton scatter compensation using the triple-energy window method for single- and dual-isotope SPECT J Nucl Med 34 2216-2221
[6]  
Ogawa K(1995)Comparative assessment of nine scatter correction methods based on spectral analysis using Monte Carlo simulations J Nucl Med 36 1476-1488
[7]  
Harata Y(1995)Multiwindow scatter correction techniques in single-photon imaging Med Phys 22 2015-2024
[8]  
Ichihara Y(1997)Scatter compensation methods in 3D iterative SPECT reconstruction: A simulation study Phys Med Biol 42 1619-1632
[9]  
Kubo A(1997)Scatter correction with an off-peak triple energy window method in Thallium-201 imaging Jpn J Mucl Med 35 789-796
[10]  
Hashimoto S(1995)A transmission-map-based scatter correction technique for SPECT in inhomogeneous media Med Phys 22 1627-1635