Determination of cocaine and metabolites in hair by column-switching LC-MS-MS analysis

被引:0
作者
Marcela Nogueira Rabelo Alves
Gabriele Zanchetti
Alberto Piccinotti
Silvia Tameni
Bruno Spinosa De Martinis
Aldo Polettini
机构
[1] Universidade de São Paulo,Faculdade de Ciências Farmacêuticas de Ribeirão Preto
[2] Synlab Italia Srl,Departamento de Química, Faculdade de Filosofia Ciências e Letras de Ribeirão Preto
[3] Universidade de São Paulo,Department of Public Health and Community Medicine
[4] University of Verona,undefined
来源
Analytical and Bioanalytical Chemistry | 2013年 / 405卷
关键词
Column switching; LC-MS-MS; Hair testing; Cocaine and metabolites; Drugs of abuse;
D O I
暂无
中图分类号
学科分类号
摘要
A method for rapid, selective, and robust determination of cocaine (CO) and metabolites in 5-mg hair samples was developed and fully validated using a column-switching liquid chromatography–tandem mass spectrometry system (LC-MS-MS). Hair samples were decontaminated, segmented, incubated overnight in diluted HCl, and centrifuged, and the diluted (1:10 with distilled water) extracts were analyzed in positive ionization mode monitoring two reactions per analyte. Quantifier transitions were: m/z 304.2→182.2 for CO, m/z 290.1→168.1 for benzoylecgonine (BE), and m/z 318.2→196.2 for cocaethylene (CE). The lower limit of quantification (LLOQ) was set at 0.05 ng/mg for CO and CE, and 0.012 ng/mg for BE. Imprecision and inaccuracy at LLOQ were lower than 20 % for all analytes. Linearity ranged between 0.05 and 50.0 ng/mg for CO and CE and 0.012 and 12.50 ng/mg for BE. Selectivity, matrix effect, process efficiency, recovery, carryover, cross talk, and autosampler stability were also evaluated during validation. Eighteen real hair samples and five samples from a commercial proficiency testing program were comparatively examined with the proposed multidimensional chromatography coupled with tandem mass spectrometry procedure and our reference gas chromatography coupled to mass spectrometry (GC-MS) method. Compared with our reference GC-MS method, column-switching technique and the high sensitivity of the tandem mass spectrometry detection system allowed to significantly reduce sample amount (×10) with increased sensitivity (×2) and sample throughput (×4), to simplify sample preparation, and to avoid that interfering compounds and ions impaired the ionization and detection of the analytes and deteriorate the performance of the ion source.
引用
收藏
页码:6299 / 6306
页数:7
相关论文
共 50 条
[31]   Simultaneous determination of deoxynivalenol, zearalenone, and their major masked metabolites in cereal-based food by LC-MS-MS [J].
Vendl, O. ;
Berthiller, F. ;
Crews, C. ;
Krska, R. .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2009, 395 (05) :1347-1354
[32]   Determination of nonsteroidal anti-inflammatory drugs in human plasma by LC-MS-MS with a hydrophilic polymer column [J].
Xiao-Pen Lee ;
Takeshi Kumazawa ;
Chika Hasegawa ;
Tetsuya Arinobu ;
Akihito Kato ;
Hiroshi Seno ;
Keizo Sato .
Forensic Toxicology, 2010, 28 :96-104
[33]   Determination of nonsteroidal anti-inflammatory drugs in human plasma by LC-MS-MS with a hydrophilic polymer column [J].
Lee, Xiao-Pen ;
Kumazawa, Takeshi ;
Hasegawa, Chika ;
Arinobu, Tetsuya ;
Kato, Akihito ;
Seno, Hiroshi ;
Sato, Keizo .
FORENSIC TOXICOLOGY, 2010, 28 (02) :96-104
[34]   Simultaneous determination of three aconitum alkaloids in urine by LC-MS-MS [J].
Wang, Zhaohong ;
Wang, Zhiping ;
Wen, Jiao ;
He, Yi .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 45 (01) :145-148
[35]   Simultaneous determination of ethyl glucuronide, cocaine, cocaethylene, and benzoylecgonine in hair by using LC-MS/MS [J].
Shin, Dong Won ;
Kim, Seon Yeong ;
Suh, Sung Ill ;
Kim, Jin Young .
JOURNAL OF ANALYTICAL SCIENCE AND TECHNOLOGY, 2024, 15 (01)
[36]   Quantitative determination of taxine B in body fluids by LC-MS-MS [J].
Frommherz, L. ;
Kintz, P. ;
Kijewski, H. ;
Koehler, H. ;
Lehr, M. ;
Brinkmann, B. ;
Beike, J. .
INTERNATIONAL JOURNAL OF LEGAL MEDICINE, 2006, 120 (06) :346-351
[37]   LC-MS-MS Determination of (S,R)-Penehyclidine in Rat Plasma [J].
Li, Changkun ;
Li, Jinglai ;
Cui, Mengxun ;
Fu, Fenghua ;
Zhong, Bohua ;
Zhang, Zhenqing .
CHROMATOGRAPHIA, 2010, 71 (11-12) :1025-1030
[38]   A sensitive and accurate quantitative method to determine N-arachidonoyldopamine and N-oleoyldopamine in the mouse striatum using column-switching LC-MS-MS: use of a surrogate matrix to quantify endogenous compounds [J].
Ji, Dajeong ;
Jang, Choon-Gon ;
Lee, Sooyeun .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2014, 406 (18) :4491-4499
[39]   LC-MS-MS Method for the Simultaneous Determination of Donepezil Enantiomers in Plasma [J].
Iordachescu, Adriana ;
Silvestro, Luigi ;
Tudoroniu, Ariana ;
Savu, Simona Rizea ;
Ciuca, Valentin .
CHROMATOGRAPHIA, 2012, 75 (15-16) :857-866
[40]   Development and Validation of a Column-Switching LC-ESI-MS Assay for Determination of Huperzine A in Rat Plasma and Cerebrospinal Fluid [J].
Pan, Jianyang ;
Wang, Qiao ;
Shao, Qing ;
Cheng, Yiyu ;
Qu, Haibin .
CHROMATOGRAPHIA, 2009, 69 (5-6) :445-451