Image quality and dose performance of 80 kV low dose scan protocol in high-pitch spiral coronary CT angiography: feasibility study

被引:0
|
作者
Dan Wang
Xiu-hua Hu
Shi-zheng Zhang
Run-ze Wu
Su-su Xie
Bin Chen
Qiao-wei Zhang
机构
[1] Zhejiang University,Department of Radiology, Sir Run Run Shaw Hospital, School of Medicine
[2] Sir Run Run Shaw Institute of Clinical Medicine of Zhejiang University,undefined
[3] Siemens Healthcare,undefined
来源
The International Journal of Cardiovascular Imaging | 2012年 / 28卷
关键词
High-pitch; Coronary CT angiography; Computed tomography; Dose; Image quality;
D O I
暂无
中图分类号
学科分类号
摘要
To investigate the image quality and dose performance of 80 kV high-pitch spiral (HPS) coronary CT angiography (CCTA). 106 patients consecutively enrolled into prospectively ECG-triggering HPS CCTA (pitch = 3.4) exam using kV/ref. mAs = 80/400, 100/370, and 120/370 when patient BMI was ≤22.5 (n = 40), between 22.5 and 27.5 (n = 53) and >27.5 kg/m² (n = 13). Image quality was assessed per-segment by two observers independently using a 4-point scale (1—excellent, 4—non-diagnosable). Image noise and signal-to-noise ratio (SNR), contrast-to-noise ratio were measured. Diagnostic image quality was obtained in 503 of 507, 687 of 693, 164 of 167 coronary segments in 80, 100, 120 kV groups without significant difference (P = 0.482). The proportions of segments with score 1–4 were not significantly different among three kV groups (all P > 0.05). Image noise were significantly higher in 80 kV group than 100 and 120 groups (P < 0.001), while SNR was not (P = 0.097). The effective dose of 80 kV group (0.36 ± 0.03 mSv) was significantly lower than that of 100 kV group (0.86 ± 0.08 mSv) and 120 kV group (1.77 ± 0.18 mSv). The mean ± SD of HR in all patients was 54.8 ± 5.1 bpm. 80 kV HPS CCTA is feasible for patient with BMI ≤ 22.5 kg/m² which can save 58% dose than 100 kV group, while maintain diagnosable image quality.
引用
收藏
页码:415 / 423
页数:8
相关论文
共 50 条
  • [31] Feasibility of Respiratory-gated High-pitch Spiral CT: Free-breathing Inspiratory Image Quality
    Lell, Michael M.
    Scharf, Michael
    Eller, Achim
    Wuest, Wolfgang
    Allmendinger, Thomas
    Fuchs, Florian
    Achenbach, Stephan
    Uder, Michael
    May, Matthias S.
    ACADEMIC RADIOLOGY, 2016, 23 (04) : 406 - 412
  • [32] Comparison of Image Quality and Radiation Dose between High-Pitch Mode and Low-Pitch Mode Spiral Chest CT in Small Uncooperative Children: The Effect of Respiratory Rate
    Kim, Seong Ho
    Choi, Young Hun
    Cho, Hyun-Hae
    Lee, So Mi
    Shin, Su-Mi
    Cheon, Jung-Eun
    Kim, Woo Sun
    Kim, In-One
    EUROPEAN RADIOLOGY, 2016, 26 (04) : 1149 - 1158
  • [33] Comparison of sequential and high-pitch-spiral coronary CT-angiography: image quality and radiation exposure
    Seppelt, D.
    Kolb, C.
    Kuehn, J. P.
    Speiser, U.
    Radosa, C. G.
    Hoberueck, S.
    Hoffmann, R. T.
    Platzek, I.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2019, 35 (07) : 1379 - 1386
  • [34] Comparison of sequential and high-pitch-spiral coronary CT-angiography: image quality and radiation exposure
    D. Seppelt
    C. Kolb
    J. P. Kühn
    U. Speiser
    C. G. Radosa
    S. Hoberück
    R. T. Hoffmann
    I. Platzek
    The International Journal of Cardiovascular Imaging, 2019, 35 : 1379 - 1386
  • [35] Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT: technique and initial experience
    Lell, Michael
    Marwan, Mohamed
    Schepis, Tiziano
    Pflederer, Tobias
    Anders, Katharina
    Flohr, Thomas
    Allmendinger, Thomas
    Kalender, Willi
    Ertel, Dirk
    Thierfelder, Carsten
    Kuettner, Axel
    Ropers, Dieter
    Daniel, Werner G.
    Achenbach, Stephan
    EUROPEAN RADIOLOGY, 2009, 19 (11) : 2576 - 2583
  • [36] High-pitch dual-source CT coronary angiography: analysis of the impact on image quality of altered electrocardiography waves during data acquisition
    Wang, Xiaoling
    Fang, Jiliang
    Tong, Haibin
    Zhao, Qing
    Song, Qingqiao
    Luo, Ping
    Xue, Chao
    Zhang, Min
    Yang, Shuhua
    Wang, Qun
    Wang, Ping
    Shi, Fengxiang
    Xu, Lei
    Rong, Peijing
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2012, 28 : 15 - 20
  • [37] Performance of turbo high-pitch dual-source CT for coronary CT angiography: first ex vivo and patient experience
    Morsbach, Fabian
    Gordic, Sonja
    Desbiolles, Lotus
    Husarik, Daniela
    Frauenfelder, Thomas
    Schmidt, Bernhard
    Allmendinger, Thomas
    Wildermuth, Simon
    Alkadhi, Hatem
    Leschka, Sebastian
    EUROPEAN RADIOLOGY, 2014, 24 (08) : 1889 - 1895
  • [38] High-Pitch Spiral Computed Tomography Effect on Image Quality and Radiation Dose in Pediatric Chest Computed Tomography
    Lell, Michael M.
    May, Matthias
    Deak, Paul
    Alibek, Sedat
    Kuefner, Michael
    Kuettner, Axel
    Koehler, Henrik
    Achenbach, Stephan
    Uder, Michael
    Radkow, Tanja
    INVESTIGATIVE RADIOLOGY, 2011, 46 (02) : 116 - 123
  • [39] High-Pitch Dual-Source Computed Tomography Renal Angiography Comparison With Conventional Low-Pitch Computed Tomography Angiography: Image Quality, Contrast Medium Volume, and Radiation Dose
    Pang, LiFang
    Zhao, Yue
    Dong, HaiPeng
    Shi, HongCheng
    Yang, WenJie
    Zhang, Huan
    Yan, FuHua
    Liu, Bo
    Yan, Jing
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2015, 39 (05) : 737 - 740
  • [40] Low-dose, 128-slice, dual-source CT coronary angiography: accuracy and radiation dose of the high-pitch and the step-and-shoot mode
    Alkadhi, Hatem
    Stolzmann, Paul
    Desbiolles, Lotus
    Baumueller, Stephan
    Goetti, Robert
    Plass, Andre
    Scheffel, Hans
    Feuchtner, Gudrun
    Falk, Volkmar
    Marincek, Borut
    Leschka, Sebastian
    HEART, 2010, 96 (12) : 933 - 938