Image quality and dose performance of 80 kV low dose scan protocol in high-pitch spiral coronary CT angiography: feasibility study

被引:0
|
作者
Dan Wang
Xiu-hua Hu
Shi-zheng Zhang
Run-ze Wu
Su-su Xie
Bin Chen
Qiao-wei Zhang
机构
[1] Zhejiang University,Department of Radiology, Sir Run Run Shaw Hospital, School of Medicine
[2] Sir Run Run Shaw Institute of Clinical Medicine of Zhejiang University,undefined
[3] Siemens Healthcare,undefined
来源
The International Journal of Cardiovascular Imaging | 2012年 / 28卷
关键词
High-pitch; Coronary CT angiography; Computed tomography; Dose; Image quality;
D O I
暂无
中图分类号
学科分类号
摘要
To investigate the image quality and dose performance of 80 kV high-pitch spiral (HPS) coronary CT angiography (CCTA). 106 patients consecutively enrolled into prospectively ECG-triggering HPS CCTA (pitch = 3.4) exam using kV/ref. mAs = 80/400, 100/370, and 120/370 when patient BMI was ≤22.5 (n = 40), between 22.5 and 27.5 (n = 53) and >27.5 kg/m² (n = 13). Image quality was assessed per-segment by two observers independently using a 4-point scale (1—excellent, 4—non-diagnosable). Image noise and signal-to-noise ratio (SNR), contrast-to-noise ratio were measured. Diagnostic image quality was obtained in 503 of 507, 687 of 693, 164 of 167 coronary segments in 80, 100, 120 kV groups without significant difference (P = 0.482). The proportions of segments with score 1–4 were not significantly different among three kV groups (all P > 0.05). Image noise were significantly higher in 80 kV group than 100 and 120 groups (P < 0.001), while SNR was not (P = 0.097). The effective dose of 80 kV group (0.36 ± 0.03 mSv) was significantly lower than that of 100 kV group (0.86 ± 0.08 mSv) and 120 kV group (1.77 ± 0.18 mSv). The mean ± SD of HR in all patients was 54.8 ± 5.1 bpm. 80 kV HPS CCTA is feasible for patient with BMI ≤ 22.5 kg/m² which can save 58% dose than 100 kV group, while maintain diagnosable image quality.
引用
收藏
页码:415 / 423
页数:8
相关论文
共 50 条
  • [21] Image quality, radiation dose, and diagnostic accuracy of prospectively ECG-triggered high-pitch coronary CT angiography at 70 kVp in a clinical setting: comparison with invasive coronary angiography
    Zhang, Long Jiang
    Wang, Yining
    Schoepf, U. Joseph
    Meinel, Felix G.
    Bayer, Richard R., II
    Qi, Li
    Cao, Jian
    Zhou, Chang Sheng
    Zhao, Yan E.
    Li, Xie
    Bin Gong, Jian
    Jin, Zhengyu
    Lu, Guang Ming
    EUROPEAN RADIOLOGY, 2016, 26 (03) : 797 - 806
  • [22] Seventy-Peak Kilovoltage High-Pitch Thoracic Aortic CT Angiography without ECG Gating: Evaluation of Image Quality and Radiation Dose
    Zhang, Long Jiang
    Zhao, Yan E.
    Schoepf, U. Joseph
    Mangold, Stefanie
    Felmly, Lloyd M.
    Li, Xie
    Tang, Chun Xiang
    Zhou, Chang Sheng
    Qi, Li
    Lu, Guang Ming
    ACADEMIC RADIOLOGY, 2015, 22 (07) : 890 - 897
  • [23] High-Pitch Coronary CT Angiography at 70 kVp Adopting a Protocol of Low Injection Speed and Low Volume of Contrast Medium
    Feng, Ruiqi
    Tong, Jiajie
    Liu, Xiaofei
    Zhao, Yu
    Zhang, Liang
    KOREAN JOURNAL OF RADIOLOGY, 2017, 18 (05) : 763 - 772
  • [24] Performance of calcium quantifications on low-dose photon-counting detector CT with high-pitch: A phantom study
    Zhou, Shanshui
    Liu, Peng
    Dong, Haipeng
    Li, Jiqiang
    Xu, Zhihan
    Schmidt, Bernhard
    Lin, Shushen
    Yang, Wenjie
    Yan, Fuhua
    Qin, Le
    HELIYON, 2024, 10 (12)
  • [25] Image quality and radiation dose of a prospectively electrocardiography-triggered high-pitch data acquisition strategy for coronary CT angiography: The multicenter, randomized PROTECTION IV study
    Deseive, Simon
    Pugliese, Francesca
    Meave, Aloha
    Alexanderson, Erick
    Martinoff, Stefan
    Hadamitzky, Martin
    Massberg, Steffen
    Hausleiter, Joerg
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2015, 9 (04) : 278 - 285
  • [26] Image quality, radiation dose, and diagnostic accuracy of prospectively ECG-triggered high-pitch coronary CT angiography at 70 kVp in a clinical setting: comparison with invasive coronary angiography
    Long Jiang Zhang
    Yining Wang
    U. Joseph Schoepf
    Felix G. Meinel
    Richard R. Bayer
    Li Qi
    Jian Cao
    Chang Sheng Zhou
    Yan E Zhao
    Xie Li
    Jian Bin Gong
    Zhengyu Jin
    Guang Ming Lu
    European Radiology, 2016, 26 : 797 - 806
  • [27] Diagnostic Performance of a Contrast-Enhanced Ultra-Low-Dose High-Pitch CT Protocol with Reduced Scan Range for Detection of Pulmonary Embolisms
    Brendlin, Andreas S.
    Winkelmann, Moritz T.
    Peisen, Felix
    Artzner, Christoph P.
    Nikolaou, Konstantin
    Othman, Ahmed E.
    Afat, Saif
    DIAGNOSTICS, 2021, 11 (07)
  • [28] Optimization of Kernel Type and Sharpness Level Improves Objective and Subjective Image Quality for High-Pitch Photon Counting Coronary CT Angiography
    Yang, Yang
    Fink, Nicola
    Emrich, Tilman
    Graafen, Dirk
    Richter, Rosa
    Bockius, Stefanie
    Wolf, Elias V.
    Laux, Gerald
    Kavermann, Larissa
    Mueller, Lukas
    Hell, Michaela
    Halfmann, Moritz C.
    DIAGNOSTICS, 2023, 13 (11)
  • [29] Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease
    Nie, Pei
    Wang, Ximing
    Cheng, Zhaoping
    Ji, Xiaopeng
    Duan, Yanhua
    Chen, Jiuhong
    EUROPEAN RADIOLOGY, 2012, 22 (10) : 2057 - 2066
  • [30] High-pitch low-voltage CT coronary artery calcium scoring with tin filtration: accuracy and radiation dose reduction
    Apfaltrer, Georg
    Albrecht, Moritz H.
    Schoepf, U. Joseph
    Duguay, Taylor M.
    De Cecco, Carlo N.
    Nance, John W.
    De Santis, Domenico
    Apfaltrer, Paul
    Eid, Marwen H.
    Eason, Chelsea D.
    Thompson, Zachary M.
    Bauer, Maximilian J.
    Varga-Szemes, Akos
    Jacobs, Brian E.
    Sorantin, Erich
    Tesche, Christian
    EUROPEAN RADIOLOGY, 2018, 28 (07) : 3097 - 3104