Accuracy and limitations of vector flow mapping: left ventricular phantom validation using stereo particle image velocimetory

被引:42
作者
Asami R. [1 ]
Tanaka T. [1 ]
Kawabata K.-I. [1 ]
Hashiba K. [1 ]
Okada T. [1 ]
Nishiyama T. [1 ]
机构
[1] Hitachi Ltd, 1-280 Higashi-Koigakubo, Kokubunji, Tokyo
关键词
Cardio-hemodynamics; Doppler ultrasound; Flow imaging; Ultrasonics;
D O I
10.1007/s12574-016-0321-5
中图分类号
学科分类号
摘要
Background: The accuracy of vector flow mapping (VFM) was investigated in comparison to stereo particle image velocimetry (stereo-PIV) measurements using a left ventricular phantom. VFM is an echocardiographic approach to visualizing two-dimensional flow dynamics by estimating the azimuthal component of flow from the mass-conservation equation. VFM provides means of visualizing cardiac flow, but there has not been a study that compared the flow estimated by VFM to the flow data acquired by other methods. Methods: A reproducible three-dimensional cardiac blood flow was created in an optically and acoustically transparent left-ventricle phantom, that allowed color-flow mapping (CFM) data and stereo-PIV to be simultaneously acquired on the same plane. A VFM algorithm was applied to the CFM data, and the resulting VFM estimation and stereo-PIV data were compared to evaluate the accuracy of VFM. Results: The velocity fields acquired by VFM and stereo-PIV were in excellent agreement in terms of the principle flow features and time-course transitions of the main vortex characteristics, i.e., the overall correlation of VFM and PIV vectors was R = 0.87 (p < 0.0001). The accuracy of VFM was suggested to be influenced by both CFM signal resolution and the three-dimensional flow, which violated the algorithm’s assumption of planar flow. Statistical analysis of the vectors revealed a standard deviation of discrepancy averaging at 4.5% over the CFM velocity range for one cardiac cycle, and that value fluctuated up to 10% depending on the phase of the cardiac cycle. Conclusions: VFM provided fairly accurate two-dimensional-flow information on cardio-hemodynamics. These findings on VFM accuracy provide the basis for VFM-based diagnosis. © 2016, The Author(s).
引用
收藏
页码:57 / 66
页数:9
相关论文
共 16 条
[1]  
Pedrizzetti G., Domenichini F., Nature optimizes the swirling flow in the human left ventricle, Phys Rev Lett, 95, (2005)
[2]  
Kheradvar A., Houle H., Pedrizzetti G., Et al., Echocardiographic particle image velocimetry: a novel technique for quantification of left ventricular blood vorticity pattern, J Am Soc Echocardiogr, 23, pp. 86-94, (2010)
[3]  
Rodriguez Munoz D., Markl M., Moya Mur J.L., Et al., Intracardiac flow visualization: current status and future directions, Eur Heart J Cardiovasc Imaging., 14, pp. 1029-1038, (2013)
[4]  
Sengupta P.P., Pedrizzetti G., Kilner P.J., Et al., Emerging trends in CV flow visualization, JACC Cardiovasc Imaging., 5, pp. 305-316, (2012)
[5]  
Ohtsuki S., Tanaka M., The flow velocity distribution from the doppler information on a plane in three-dimensional flow, J Vis., 9, pp. 69-82, (2006)
[6]  
Garcia D., Del Alamo J.C., Tanne D., Et al., Two-dimensional intraventricular flow mapping by digital processing conventional color-Doppler echocardiography images, IEEE Trans Med Imaging, 29, pp. 1701-1713, (2010)
[7]  
Itatani K., Okada T., Uejima T., Et al., Intraventricular flow velocity vector visualization based on the continuity equation and measurements of vorticity and wall shear stress, Jpn J Appl Phys., 52, (2013)
[8]  
Uejima T., Koike A., Sawada H., Et al., A new echocardiographic method for identifying vortex flow in the left ventricle: numerical validation, Ultrasound Med Biol, 36, pp. 772-788, (2010)
[9]  
Egorov V., Tsyuryupa S., Kanilo S., Et al., Soft tissue elastometer, Med Eng Phys, 30, pp. 206-212, (2008)
[10]  
Kusunose K., Chono T., Tabata T., Et al., Echocardiographic image tracker with a speckle adaptive noise reduction filter for the automatic measurement of the left atrial volume curve, Eur Heart J Cardiovasc Imaging., 15, pp. 509-514, (2014)