Airway troubles related to the double-lumen endobronchial tube in thoracic surgery

被引:4
作者
Taguchi H. [1 ]
Yamada K. [1 ]
Matsumoto H. [1 ]
Kato A. [1 ]
Imanishi T. [1 ]
Shingu K. [1 ]
机构
[1] Department of Anesthesiology, Kansai Medical University, Moriguchi, Osaka 570
关键词
Airway obstruction; Complications; Double-lumen tube; One-lung ventilation; Tube troubles;
D O I
10.1007/BF02480033
中图分类号
学科分类号
摘要
Purpose. Several case reports indicate critical respiratory complications in relation to the double-lumen endobronchial tube (DLT). A prospective survey for the airway problems in using the DLT is presented. Methods. One hundred adult patients undergoing thoracotomy for lung cancer were investigated. Tube malposition and airway obstruction were searched using a fiber-optic scope. The endobronchial cuff was positioned just below the tracheal carina while the trachea was intubated with a DLT (Rusch). The distances of displacement, from the tracheal carina to the bronchial cuff, were measured during anesthesia using an epidural catheter, which had marks every 5 mm. The distances for correcting the tube position were measured at both the bronchial cuff and the level of the teeth. PaO2, PaCO2 and SPO2 were also measured. Results. Malposition (displacement over 5 mm from the correct position) was found in 42 patients, and 40 of them were in a withdrawal direction, occurring at the postural change and during one-lung ventilation, especially during manipulation of the lung hilum. Correcting distances at the level of the teeth were 1.5-3-times longer than those at the bronchial cuff. Airway deformities and gradual withdrawal of the bronchial cuff were found in association with surgical manipulation. Obstruction occurred at the tips of the tracheal tube in four patients and the bronchial tube in six patients, and at the tip of both in two patients. Hypoxemia (PaO2 < 60 mmHg) occurred in four patients and hypercapnea (PaCO2 > 60 mmHg) in two patients. Conclusion. Most of the DLT obstructions were associated with withdrawal malposition. Great attention to DLT displacement and airway deformity is advised.
引用
收藏
页码:173 / 178
页数:5
相关论文
共 36 条
[1]  
Maguire DP(1987)Bronchial obstruction and hypoxia during one-lung ventilation Anesthesiology 66 830-831
[2]  
Spiro AW(1987)Cardiac arrest from pulmonary outflow tract obstruction due to a double-lumen tube Anesthesiology 66 422-423
[3]  
Wells DG(1994)Robertshow double lumen tubes: a reappraisal thirty years on Anaesth Intens Care 22 179-183
[4]  
Zelcer J(1993)A quality improvement study of the placement and complications of double-lumen endobronchial tubes Cardiothorac Vasc Anesth 7 517-520
[5]  
Podolakin W(1992)The utility of a double-lumen tube for one-lung ventilation in a variety of noncardiac thoracic surgical procedures Cardiothorac Vasc Anesth 6 705-710
[6]  
Baker TG(1976)Alteration of endotracheal tube position: flexion and extension of the neck Crit Care Med 4 8-12
[7]  
Wilson AC(1985)Alteration of double-lumen endobronchial tube position by flexion and extension of the neck Anesthesiology 62 696-697
[8]  
White AL(1984)Anaesthesia and mediastinal masses. A case report and review of the literature Anaesthesia 39 899-903
[9]  
Conacher ID(1975)Anatomical factors in difficult direct laryngoscopy Br J Anaesth 47 468-474
[10]  
Herrema IH(1990)Double-lumen tube design fault Anesthesiology 73 575-576