Bioanalytical Method for the Determination of Hydroxyproline in Mouse Kidney by High-Performance Liquid Chromatography with Tandem Mass Spectrometric Detection

被引:0
作者
Rika Shiota
Hirotoshi Morita
Tomoko Matsumoto
Atsushi Morimoto
Jun Hayakawa
Masako Oka
Hiroshi Kamimori
机构
[1] Shionogi TechnoAdvance Research Co.,Pharmaceutical Research Division
[2] Ltd.,undefined
[3] Biomarker R&D Department,undefined
[4] Shionogi & Co. Ltd.,undefined
[5] Shionogi & Co. Ltd,undefined
来源
Analytical Sciences | 2017年 / 33卷
关键词
Hydroxyproline; mass spectrometry; mouse; kidney;
D O I
暂无
中图分类号
学科分类号
摘要
An analytical method was developed and validated for the measurement of hydroxyproline (Hyp) levels in mouse kidney by high-performance liquid chromatography with tandem mass spectrometric detection (LC/MS/MS) using an analytical column specially designed for the LC/MS/MS analysis for intact amino acids. Tissue hydrolyzed with hydrochloric acid could be directly injected into the LC/MS/MS, as well as separated and detected using the deuterium labelled Hyp as an internal standard. The calibration curve showed good linearity from 5 to 500 nmol/mg of tissue; the precision and accuracy, including within- and between-run, were less than 6% and within 100 ± 6%, respectively.
引用
收藏
页码:719 / 722
页数:3
相关论文
共 101 条
[1]  
Parviainen M T(1999)undefined Clin. Chim. Acta 279 145-undefined
[2]  
Jaaskelainen K(2002)undefined Toxicol. Lett. 134 319-undefined
[3]  
Kroger H(2009)undefined Kidney Int. 75 1145-undefined
[4]  
Arnala I(2010)undefined Drug Discov. Today Dis. Models 7 13-undefined
[5]  
Alhava E(2015)undefined Biomed. Pharmacoter. 70 174-undefined
[6]  
Perdelli F(2001)undefined J. Chromatogr. B 757 369-undefined
[7]  
Cristina M L(2003)undefined J. Chromatogr. B 791 427-undefined
[8]  
Sartini M(2003)undefined J. Pharm. Biomed. Anal. 33 1081-undefined
[9]  
Orlando P(2007)undefined J. Chromatogr. B 860 78-undefined
[10]  
Chevalier R L(2007)undefined J. Chromatogr. A 1212 150-undefined