Impact of gap size and groove design of hydrodynamic bearing on plasma skimming effect for use in rotary blood pump

被引:0
作者
Ming Jiang
Daisuke Sakota
Ryo Kosaka
Wataru Hijikata
机构
[1] Tokyo Institute of Technology,Department of Mechanical Engineering, School of Engineering
[2] National Institute of Advanced Industrial Science and Technology (AIST),Artificial Organ Research Group, Health and Medical Research Institute
来源
Journal of Artificial Organs | 2022年 / 25卷
关键词
Hemolysis; Hydrodynamic bearing; Plasma skimming; Rotary blood pump; Spiral groove pattern;
D O I
暂无
中图分类号
学科分类号
摘要
Plasma skimming can exclude red blood cells from high shear regions in the gaps formed by hydrodynamic bearings in rotary blood pumps. We investigated the effect of the gap size and groove design on the plasma skimming efficiency. Spiral groove bearings (SGBs) were installed into a specially designed test rig for in vitro experiments performed using human blood. The measured gap between the ridges of the bearing and the rotor surface was 17–26 µm at a flow rate of 150 ml/min and a rotor speed of 2400 rpm. Three different patterns of SGBs were designed (SGB-0, SGB-30, and SGB-60) with various degrees of the circumferential component. The hematocrit measured by a high-speed camera was compared with the hematocrit in the circuit, and the plasma skimming efficiency for the three bearing patterns was evaluated at hematocrits of 20%, 25%, and 30%. SGB-60, which had the strongest circumferential component, provided the best plasma skimming efficiency. When the gap size was less than 20 µm, the red blood cells in the gaps between the ridges of the bearing and rotor surface reduced significantly and the efficiency became higher than 90%. The gap size had the strongest effect on producing a significant plasma skimming. The plasma skimming efficiency can be significantly improved by optimizing the bearing gap size and groove design, which facilitates the further development of SGBs for use in applications such as rotary blood pumps.
引用
收藏
页码:195 / 203
页数:8
相关论文
共 13 条
  • [1] Impact of gap size and groove design of hydrodynamic bearing on plasma skimming effect for use in rotary blood pump
    Jiang, Ming
    Sakota, Daisuke
    Kosaka, Ryo
    Hijikata, Wataru
    JOURNAL OF ARTIFICIAL ORGANS, 2022, 25 (03) : 195 - 203
  • [2] Spiral groove bearing design for improving plasma skimming in rotary blood pumps
    Jiang, Ming
    Hijikata, Wataru
    JOURNAL OF ARTIFICIAL ORGANS, 2024, 27 (03) : 212 - 221
  • [3] Plasma skimming efficiency of human blood in the spiral groove bearing of a centrifugal blood pump
    Sakota, Daisuke
    Kondo, Kazuki
    Kosaka, Ryo
    Nishida, Masahiro
    Maruyama, Osamu
    JOURNAL OF ARTIFICIAL ORGANS, 2021, 24 (02) : 126 - 134
  • [4] Plasma skimming efficiency of human blood in the spiral groove bearing of a centrifugal blood pump
    Daisuke Sakota
    Kazuki Kondo
    Ryo Kosaka
    Masahiro Nishida
    Osamu Maruyama
    Journal of Artificial Organs, 2021, 24 : 126 - 134
  • [5] Plasma Skimming in a Spiral Groove Bearing of a Centrifugal Blood Pump
    Murashige, Tomotaka
    Sakota, Daisuke
    Kosaka, Ryo
    Nishida, Masahiro
    Kawaguchi, Yasuo
    Yamane, Takashi
    Maruyama, Osamu
    ARTIFICIAL ORGANS, 2016, 40 (09) : 856 - 866
  • [6] The Spiral Groove Bearing as a Mechanism for Enhancing the Secondary Flow in a Centrifugal Rotary Blood Pump
    Amaral, Felipe
    Gross-Hardt, Sascha
    Timms, Daniel
    Egger, Christina
    Steinseifer, Ulrich
    Schmitz-Rode, Thomas
    ARTIFICIAL ORGANS, 2013, 37 (10) : 866 - 874
  • [7] Optimal Design of the Hydrodynamic Multi-Arc Bearing in a Centrifugal Blood Pump for the Improvement of Bearing Stiffness and Hemolysis Level
    Yasui, Kazuya
    Kosaka, Ryo
    Nishida, Masahiro
    Maruyama, Osamu
    Kawaguchi, Yasuo
    Yamane, Takashi
    ARTIFICIAL ORGANS, 2013, 37 (09) : 768 - 777
  • [8] A hydrodynamically suspended, magnetically sealed mechanically noncontact axial flow blood pump: Design of a hydrodynamic bearing
    Mitamura, Yoshinori
    Kido, Kazuyuki
    Yano, Tetsuya
    Sakota, Daisuke
    Yambe, Tomoyuki
    Sekine, Kazumitsu
    OKamoto, Eiji
    ARTIFICIAL ORGANS, 2007, 31 (03) : 221 - 224
  • [9] Investigation of the Axial Gap Clearance in a Hydrodynamic-Passive Magnetically Levitated Rotary Blood Pump Using X-Ray Radiography
    Thamsen, Bente
    Plamondon, Mathieu
    Granegger, Marcus
    Daners, Marianne Schmid
    Kaufmann, Rolf
    Neels, Antonia
    Meboldt, Mirko
    ARTIFICIAL ORGANS, 2018, 42 (05) : 510 - 515
  • [10] Effect of a bearing gap on hemolytic property in a hydrodynamically levitated centrifugal blood pump with a semi-open impeller
    Kosaka, Ryo
    Nishida, Masahiro
    Maruyama, Osamu
    Yambe, Tomoyuki
    Imachi, Kou
    Yamane, Takashi
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2013, 23 (1-2) : 37 - 47