Influence of cup-center-edge angle on micro-motion at the interface between the cup and host bone in cementless total hip arthroplasty: three-dimensional finite element analysis

被引:7
作者
Kaku N. [1 ]
Tabata T. [1 ]
Tsumura H. [1 ]
机构
[1] Department of Orthopaedic Surgery, Faculty of Medicine, Oita University, 1-1 Idaigaoka Hazamacho, Yufu City, 879-5593, Oita
关键词
Bone graft; Cementless; Cup-center-edge angle; Dysplasia; Finite element model; Micro-motion;
D O I
10.1007/s00590-015-1697-z
中图分类号
学科分类号
摘要
We verified the index cup position required for bulk bone grafting instead of morcellized grafting immediately after cementless total hip arthroplasty. Three-dimensional finite element analysis was used to evaluate changes in the volume of the slippage of the cup–host bone interface as micro-motion of the cup at the acetabular bone defect site depending on the cup-center-edge (CE) angle. The conditions of bulk bone grafts were similar to those of cortical bone. Slippage increased with decreasing cup-CE angle. A bulk bone graft tightly fixed to the host bone prevented considerably larger slippage between the cup and host bone. A smaller cup-CE angle increased the impact of the bulk bone graft on slippage. When the cup-CE angle was 0° or −10°, the criterion for slippage in favorable initial fixation in all conditions was <40 μm. Even if transplanted bulk bone is used, unless good fixation is obtained between the host bone, and the cup and bone graft, it is impossible to obtain reliable fixation of the cup with a cup-CE angle <−10° and slippage exceeding 40 μm. Bulk bone grafting tightly fixed to the host bone improves initial the cup–host bone fixation, especially when the cup-CE angle is small, such as <−10°. In clinical practice, negative factors are implicated in the initial fixation of various cups, and sufficient fixation between the host bone and cup or bulk bone graft using a screw is effective when the cup-CE angle is extremely small. © 2015, Springer-Verlag France.
引用
收藏
页码:1271 / 1277
页数:6
相关论文
共 30 条
[1]  
Johnston R.C., Brand R.A., Crowninshield R.D., Reconstruction of the hip. A mathematical approach to determine optimum geometric relationships, J Bone Joint Surg, 61, 5, pp. 639-652, (1979)
[2]  
Delp S.L., Wixson R.L., Komattu A.V., Kocmond J.H., How superior placement of the joint center in hip arthroplasty affects the abductor muscles, Clin Orthop Relat Res, 328, pp. 137-146, (1996)
[3]  
Jerosch J., Steinbeck J., Stechmann J., Guth V., Influence of a high hip center on abductor muscle function, Arch Orthop Trauma Surg, 116, 6-7, pp. 385-389, (1997)
[4]  
Goto K., Akiyama H., Kawanabe K., So K., Morimoto T., Nakamura T., Long-term results of cemented total hip arthroplasty for dysplasia, with structural autograft fixed with poly-l-lactic acid screws, J Arthroplasty, 24, 8, pp. 1146-1151, (2009)
[5]  
Saito S., Ishii T., Mori S., Hosaka K., Nemoto N., Tokuhashi Y., Long-term results of bulk femoral head autograft in cementless THA for developmental hip dysplasia, Orthopedics, 34, 2, (2011)
[6]  
Tsukada S., Wakui M., Bulk femoral head autograft without decortication in uncemented total hip arthroplasty: seven-to ten-year results, J Arthroplasty, 27, 3, pp. 437-444, (2012)
[7]  
Sugano N., Nishii T., Nakata K., Masuhara K., Takaoka K., Polyethylene sockets and alumina ceramic heads in cemented total hip arthroplasty. A ten-year study, J Bone Joint Surg Br, 77, 4, pp. 548-556, (1995)
[8]  
Burstein A.H., Reilly D.T., Martens M., Aging of bone tissue: mechanical properties, J Bone Joint Surg Am, 58, 1, pp. 82-86, (1976)
[9]  
Wirtz D.C., Schiffers N., Pandorf T., Radermacher K., Weichert D., Forst R., Critical evaluation of known bone material properties to realize anisotropic FE-simulation of the proximal femur, J Biomech, 33, 10, pp. 1325-1330, (2000)
[10]  
Heller M.O., Bergmann G., Kassi J.P., Claes L., Haas N.P., Duda G.N., Determination of muscle loading at the hip joint for use in pre-clinical testing, J Biomech, 38, 5, pp. 1155-1163, (2005)