Biomechanical properties of five different currently used implants for open-wedge high tibial osteotomy

被引:42
作者
Diffo Kaze A. [1 ]
Maas S. [1 ]
Waldmann D. [1 ]
Zilian A. [1 ]
机构
[1] Technology and Communication, University of Luxembourg, 6, rue R. Coudenhove-Kalergi, Luxembourg
关键词
Biomechanics; ContourLock; Cyclic and static loading; High tibial osteotomy (HTO); iBalance; Mechanical test; Osteoarthritis; PEEKPower; Stiffness; TomoFix; Valgus malrotation;
D O I
10.1186/s40634-015-0030-4
中图分类号
学科分类号
摘要
Background: As several new tibial osteotomy plates recently appeared on the market, the aim of the present study was to compare mechanical static and fatigue strength of three newly designed plates with gold standard plates for the treatment of medial knee joint osteoarthritis. Methods: Sixteen fourth-generation tibial bone composites underwent a medial open-wedge high tibial osteotomy (HTO) according to standard techniques, using five TomoFix standard plates, five PEEKPower plates and six iBalance implants. Static compression load to failure and load-controlled cyclic fatigue failure tests were performed. Forces, horizontal and vertical displacements were measured; rotational permanent plastic deformations, maximal displacement ranges in the hysteresis loops of the cyclic loading responses and dynamic stiffness were determined. Results: Static compression load to failure tests revealed that all plates showed sufficient stability up to 2400 N without any signs of opposite cortex fracture, which occurred above this load in all constructs at different load levels. During the fatigue failure tests, screw breakage in the iBalance group and opposite cortex fractures in all constructs occurred only under physiological loading conditions (<2400 N). The highest fatigue strength in terms of maximal load and number of cycles performed prior to failure was observed for the ContourLock group followed by the iBalance implants, the TomoFix standard (std) and small stature (sm) plates. The PEEKPower group showed the lowest fatigue strength. Conclusions: All plates showed sufficient stability under static loading. Compared to the TomoFix and the PEEKPower plates, the ContourLock plate and iBalance implant showed a higher mechanical fatigue strength during cyclic fatigue testing. These data suggest that both mechanical static and fatigue strength increase with a wider proximal T-shaped plate design together with diverging proximal screws as used in the ContourLock plate or a closed-wedge construction as in the iBalance design. Mechanical strength of the bone-implant constructs decreases with a narrow T-shaped proximal end design and converging proximal screws (TomoFix) or a short vertical plate design (PEEKPower Plate). Whenever high mechanical strength is required, a ContourLock or iBalance plate should be selected. © 2015, Diffo Kaze et al.
引用
收藏
页码:1 / 17
页数:16
相关论文
共 30 条
[1]  
Agneskirchner J., Freiling D., Hurschler C., Lobenhoffer P., Primary stability of four different implants for opening wedge high tibial osteotomy, Knee Surg Sports Traumatol Arthrosc, 14, pp. 291-300, (2006)
[2]  
Brinkman J.-M., Hurschler C., Agneskirchner J., Lobenhoffer P., Castelein R.M., Van Heerwaarden R.J., Biomechanical testing of distal femur osteotomy plate fixation techniques: The role of simulated physiological loading, J Exp Orthop, 1, (2014)
[3]  
Chong A.C., Miller F., Buxton M., Friis E.A., Fracture Toughness and fatigue crack propagation rate of short fiber reinforced epoxy composites for analogue cortical bone, J Biomech Eng, 129, 4, pp. 487-493, (2007)
[4]  
Cotic M., Vogt S., Hinterwimmer S., Feucht M.J., Slotta-Huspenina J., Schuster T., Imhoff A.B., A matched-pair comparison of two different locking plates for valgus-producing medial open-wedge high tibial osteotomy: peek-carbon composite plate versus titanium plate, Knee Surgery, (2014)
[5]  
Fowler P.J., Tan J.L., Brown G.A., Medial opening wedge high tibial osteotomy: how I do it, Oper Tech Sports Med, 8, pp. 32-38, (2000)
[6]  
Gardner M.P., Chong A.C.M., Pollock A.G., Wooley P.H., Mechanical Evaluation of Large-Size Fourth-Generation Composite Femur and Tibia Models, Ann Biomed Eng, 38, 3, pp. 613-620, (2010)
[7]  
Heiner A.D., Structural properties of fourth-generation composite femurs and tibias, J Biomech, 41, pp. 3282-3284, (2008)
[8]  
Heinlein B., Kutzner I., Graichen F., Bender A., Rohlmann A., Halder A.M., Beier A., Bergmann G., ESB clinical biomechanics award 2008: Complete data of total knee replacement loading for level walking and stair climbing measured in vivo with a follow-up of 6-101 months, Clin Biomech, 24, pp. 315-326, (2009)
[9]  
Implants for surgery: Determination of endurance properties of stemmed femoral components with application of torsion, (1989)
[10]  
Implants for surgery: Determination of endurance properties of head and neck region of stemmed femoral components, (1992)