Measuring Systems for Thermometer Calibration in Low-Temperature Range

被引:0
作者
A. Szmyrka-Grzebyk
L. Lipiński
H. Manuszkiewicz
A. Kowal
A. Grykałowska
D. Jancewicz
机构
[1] Institute of Low Temperature and Structure Research,
来源
International Journal of Thermophysics | 2011年 / 32卷
关键词
Calibration; Temperature scale; Thermometry;
D O I
暂无
中图分类号
学科分类号
摘要
The national temperature standard for the low-temperature range between 13.8033 K and 273.16 K has been established in Poland at the Institute of Low Temperature and Structure Research (INTiBS). The standard consists of sealed cells for realization of six fixed points of the International Temperature Scale of 1990 (ITS-90) in the low-temperature range, an adiabatic cryostat and Isotech water and mercury triple-point baths, capsule standard resistance thermometers (CSPRT), and AC and DC bridges with standard resistors for thermometers resistance measurements. INTiBS calibrates CSPRTs at the low-temperature fixed points with uncertainties less than 1 mK. In lower temperature range—between 2.5 K and about 25 K — rhodium–iron (RhFe) resistance thermometers are calibrated by comparison with a standard which participated in the EURAMET.T-K1.1 comparison. INTiBS offers a calibration service for industrial platinum resistance thermometers and for digital thermometers between 77 K and 273 K. These types of thermometers may be calibrated at INTiBS also in a higher temperature range up to 550°C. The Laboratory of Temperature Standard at INTiBS acquired an accreditation from the Polish Centre for Accreditation. A management system according to EN ISO/IEC 17025:2005 was established at the Laboratory and presented on EURAMET QSM Forum.
引用
收藏
页码:2466 / 2476
页数:10
相关论文
共 50 条
[41]   Uncertainty evaluation of proposed setup for the calibration of vertical angle measuring systems by using means for the flat angle calibration [J].
Siaudinyte, Lauryna ;
Suh, Ho Suhng .
MEASUREMENT, 2015, 67 :177-182
[42]   Low-Cost Calibration Techniques for Smart Temperature Sensors [J].
Pertijs, Michiel A. P. ;
Aita, Andre L. ;
Makinwa, Kofi A. A. ;
Huijsing, Johan H. .
IEEE SENSORS JOURNAL, 2010, 10 (06) :1098-1105
[43]   A novel robot-assisted calibration procedure for Optical Coordinate Measuring Systems [J].
Gill, Connor ;
Haynes, Adam ;
Justham, Laura ;
Dobrzanski, James ;
Buswell, Richard ;
Lohse, Niels ;
Kinnell, Peter .
PRECISION ENGINEERING-JOURNAL OF THE INTERNATIONAL SOCIETIES FOR PRECISION ENGINEERING AND NANOTECHNOLOGY, 2025, 96 :55-64
[44]   Calibration of cameras of photogrammetric measuring systems with the use of a genetic solution search algorithm [J].
B. N. Markov ;
S. G. Konov ;
A. A. Loginov .
Measurement Techniques, 2012, 55 :551-554
[45]   Calibration of cameras of photogrammetric measuring systems with the use of a genetic solution search algorithm [J].
Markov, B. N. ;
Konov, S. G. ;
Loginov, A. A. .
MEASUREMENT TECHNIQUES, 2012, 55 (05) :551-554
[46]   Wide-Aperture, Low-Temperature Noise Generators for Calibration of the Radio Telescopes of Two-Element Radio Interferometer with Very Long Baseline [J].
Arsaev, I. E. ;
Lapshin, A. I. ;
Sargsyan, M. V. ;
Yurchuk, E. F. .
MEASUREMENT TECHNIQUES, 2018, 61 (03) :290-296
[47]   Ruby pressure scale in a low-temperature diamond anvil cell [J].
Yamaoka, Hitoshi ;
Zekko, Yumiko ;
Jarrige, Ignace ;
Lin, Jung-Fu ;
Hiraoka, Nozomu ;
Ishii, Hirofumi ;
Tsuei, Ku-Ding ;
Mizuki, Jun'ichiro .
JOURNAL OF APPLIED PHYSICS, 2012, 112 (12)
[48]   On the SiO2-OH capacitance low-temperature thermometers [J].
Rotter, M ;
Boutard, D ;
Ladieu, F ;
Pari, P .
PHYSICA B, 2000, 284 :1994-1995
[49]   Calibration of Radiation Thermometers and Blackbodies in the Temperature Range from 160 °C to 420 °C [J].
Y. Shimizu ;
J. Ishii .
International Journal of Thermophysics, 2011, 32 :2560-2578
[50]   Calibration of Radiation Thermometers and Blackbodies in the Temperature Range from 160 °C to 420 °C [J].
Shimizu, Y. ;
Ishii, J. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2011, 32 (11-12) :2560-2578