Measuring Systems for Thermometer Calibration in Low-Temperature Range

被引:0
|
作者
A. Szmyrka-Grzebyk
L. Lipiński
H. Manuszkiewicz
A. Kowal
A. Grykałowska
D. Jancewicz
机构
[1] Institute of Low Temperature and Structure Research,
来源
关键词
Calibration; Temperature scale; Thermometry;
D O I
暂无
中图分类号
学科分类号
摘要
The national temperature standard for the low-temperature range between 13.8033 K and 273.16 K has been established in Poland at the Institute of Low Temperature and Structure Research (INTiBS). The standard consists of sealed cells for realization of six fixed points of the International Temperature Scale of 1990 (ITS-90) in the low-temperature range, an adiabatic cryostat and Isotech water and mercury triple-point baths, capsule standard resistance thermometers (CSPRT), and AC and DC bridges with standard resistors for thermometers resistance measurements. INTiBS calibrates CSPRTs at the low-temperature fixed points with uncertainties less than 1 mK. In lower temperature range—between 2.5 K and about 25 K — rhodium–iron (RhFe) resistance thermometers are calibrated by comparison with a standard which participated in the EURAMET.T-K1.1 comparison. INTiBS offers a calibration service for industrial platinum resistance thermometers and for digital thermometers between 77 K and 273 K. These types of thermometers may be calibrated at INTiBS also in a higher temperature range up to 550°C. The Laboratory of Temperature Standard at INTiBS acquired an accreditation from the Polish Centre for Accreditation. A management system according to EN ISO/IEC 17025:2005 was established at the Laboratory and presented on EURAMET QSM Forum.
引用
收藏
页码:2466 / 2476
页数:10
相关论文
共 50 条
  • [31] Miniature high-resolution thermometer for low-temperature applications
    Welander, PB
    Hahn, I
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (09): : 3600 - 3604
  • [32] ELECTROACOUSTICAL THERMOMETRY IN LOW-TEMPERATURE RANGE
    BRODSKII, AD
    MEASUREMENT TECHNIQUES-USSR, 1967, (06): : 671 - &
  • [33] MAGNETIC REFRIGERATION IN THE LOW-TEMPERATURE RANGE
    TISHIN, AM
    JOURNAL OF APPLIED PHYSICS, 1990, 68 (12) : 6480 - 6484
  • [34] THERMAL ANCHORING AND CALIBRATION OF LOW-TEMPERATURE SENSORS
    KOEPPE, W
    KERNTECHNIK, 1973, 15 (08) : 353 - 357
  • [35] Low-temperature blackbodies and facility for calibration of them
    Morozova, SP
    Morozov, PA
    Sapritsky, VI
    Lisiansky, BE
    Dovgilov, NL
    Liang, YX
    Li, YM
    Yue, WL
    OPTICAL DESIGN AND TESTING, 2002, 4927 : 118 - 124
  • [36] ON MEASURING THE TEMPERATURE OF LIQUID STEEL WITH LOW-TEMPERATURE METERS
    KARPOV, IV
    INDUSTRIAL LABORATORY, 1960, 26 (06): : 782 - 785
  • [37] AGING PHENOMENA IN A LOW-TEMPERATURE GLASS-CERAMIC CAPACITANCE THERMOMETER
    LAWLESS, WN
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1975, 46 (05): : 625 - 628
  • [38] STANDARD LOW-TEMPERATURE VIBRATION-PROOF PLATINUM RESISTANCE THERMOMETER
    VORFOLOMEEV, SF
    KYTIN, GA
    ANOKHINA, TN
    MEASUREMENT TECHNIQUES, 1980, 23 (02) : 125 - 125
  • [39] A SQUID-based primary noise thermometer for low-temperature metrology
    Kirste, A.
    Engert, J.
    PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2016, 374 (2064):
  • [40] Extremely Low-Temperature Systems
    Easterbrook, Nancy
    Boland, Tim
    Farese, Dave
    CHEMICAL ENGINEERING, 2015, 122 (08) : 38 - 44