Finite element model of the impaction of a press-fitted acetabular cup

被引:0
|
作者
Adrien Michel
Vu-Hieu Nguyen
Romain Bosc
Romain Vayron
Philippe Hernigou
Salah Naili
Guillaume Haiat
机构
[1] UMR CNRS 8208,Université Paris Est, Laboratoire de Modélisation et de Simulation Multi
[2] Hôpital Henri Mondor AP-HP,Echelle
[3] CHU Paris 12,Service de Chirurgie Plastique et Reconstructive
[4] Université Paris-Est,Service de Chirurgie Orthopédique
[5] Hôpital Henri Mondor AP-HP,CNRS, Laboratoire de Modélisation et de Simulation Multi‑Echelle
[6] CHU Paris 12,Research Center
[7] Université Paris-Est,undefined
[8] UMR CNRS 8208,undefined
[9] École de technologie supérieure,undefined
[10] Hôpital du Sacré-Cœur de Montréal,undefined
来源
Medical & Biological Engineering & Computing | 2017年 / 55卷
关键词
Acetabular cup; Hip replacement; Impact; Finite element; Bone; Biomechanics;
D O I
暂无
中图分类号
学科分类号
摘要
Press-fit surgical procedures aim at providing primary stability to acetabular cup (AC) implants. Impact analysis constitutes a powerful approach to retrieve the AC implant insertion properties. The aim of this numerical study was to investigate the dynamic interaction occurring between the hammer, the ancillary and bone tissue during the impact and to assess the potential of impact analysis to retrieve AC implant insertion conditions. A dynamic two-dimensional axisymmetric model was developed to simulate the impaction of the AC implant into bone tissue assuming friction at the bone–implant interface and large deformations. Different values of interference fit (from 0.5 to 2 mm) and impact velocities (from 1 to 2 m.s−1) were considered. For each configuration, the variation of the force applied between the hammer and the ancillary was analyzed and an indicator I was determined based on the impact momentum of the signal. The simulated results are compared to the experiments. The value of the polar gap decreases with the impact velocity and increases with the interference fit. The bone–implant contact area was significantly correlated with the resonance frequency (R2 = 0.94) and the indicator (R2 = 0.95). The results show the potential of impact analyses to retrieve the bone–implant contact properties.
引用
收藏
页码:781 / 791
页数:10
相关论文
共 50 条
  • [1] Finite element model of the impaction of a press-fitted acetabular cup
    Michel, Adrien
    Vu-Hieu Nguyen
    Bosc, Romain
    Vayron, Romain
    Hernigou, Philippe
    Naili, Salah
    Haiat, Guillaume
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2017, 55 (05) : 781 - 791
  • [2] 3-D finite element model of the impaction of a press-fitted femoral stem under various biomechanical environments
    Poudrel, Anne-Sophie
    Bouffandeau, Arthur
    Rosi, Giuseppe
    Dubory, Arnaud
    Lachaniette, Charles-Henri Flouzat
    Nguyen, Vu-Hieu
    Haiat, Guillaume
    Computers in Biology and Medicine, 2024, 174
  • [3] Damage Analysis and Finite Element Simulation of Fretting Wear on Press-Fitted Surface of Railway Axle
    Yuan C.
    Li Y.
    Yang K.
    Lu L.
    Zou L.
    Li W.
    Zeng D.
    Mocaxue Xuebao/Tribology, 2020, 40 (04): : 520 - 530
  • [4] Fretting Wear Simulation of Press-Fitted Shaft with Finite Element Analysis and Influence Function Method
    Lee, Dong-Hyong
    Kwon, Seok-Jin
    Choi, Jae-Boong
    Kim, Young-Jin
    TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS A, 2008, 32 (01) : 54 - 62
  • [5] Deformation of press-fitted metallic resurfacing cups. Part 2: finite element simulation
    Yew, A
    Jin, ZM
    Donn, A
    Morlock, MM
    Isaac, G
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) : 311 - 319
  • [6] Explicit finite element modelling of the impaction of metal press-fit acetabular components
    Hothi, H. S.
    Busfield, J. J. C.
    Shelton, J. C.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2011, 225 (H3) : 303 - 314
  • [7] Finite Element Modeling and Experimental Validation of Fretting Wear Scars in a Press-Fitted Shaft with an Open Zone
    Zhang, Yuanbin
    Lu, Liantao
    Gong, Yubin
    Zhang, Jiwang
    Zeng, Dongfang
    TRIBOLOGY TRANSACTIONS, 2018, 61 (04) : 585 - 595
  • [8] SHAPE TOLERANCES FOR PRESS-FITTED COMPONENTS
    DMITRIEV, AM
    RUSSIAN ENGINEERING JOURNAL-USSR, 1965, 45 (10): : 43 - &
  • [9] Primary stability of a press-fit cup in combination with impaction grafting in an acetabular defect model
    Schierjott, Ronja A.
    Hettich, Georg
    Baxmann, Marc
    Morosato, Federico
    Cristofolini, Luca
    Grupp, Thomas M.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2021, 39 (05) : 929 - 940
  • [10] Pulling apart a press-fitted joint