Patient-specific in silico endovascular repair of abdominal aortic aneurysms: application and validation

被引:0
作者
André Hemmler
Brigitta Lutz
Günay Kalender
Christian Reeps
Michael W. Gee
机构
[1] Technische Universität München,Mechanics and High Performance Computing Group
[2] Universitätsklinikum Carl Gustav Carus Dresden,undefined
[3] Klinik für Viszeral-,undefined
[4] Thorax- und Gefäßchirurgie,undefined
[5] DRK Kliniken Berlin,undefined
[6] Klinik für vaskuläre und endovaskuläre Chirurgie,undefined
来源
Biomechanics and Modeling in Mechanobiology | 2019年 / 18卷
关键词
Abdominal aortic aneurysm; Endovascular repair; Stent-graft; Patient-specific modeling; Finite element method;
D O I
暂无
中图分类号
学科分类号
摘要
Non-negligible postinterventional complication rates after endovascular aneurysm repair (EVAR) leave room for further improvements. Since the potential success of EVAR depends on various patient-specific factors, such as the complexity of the vessel geometry and the physiological state of the vessel, in silico models can be a valuable tool in the preinterventional planning phase. A suitable in silico EVAR methodology applied to patient-specific cases can be used to predict stent-graft (SG)-related complications, such as SG migration, endoleaks or tissue remodeling-induced aortic neck dilatation and to improve the selection and sizing process of SGs. In this contribution, we apply an in silico EVAR methodology that predicts the final state of the deployed SG after intervention to three clinical cases. A novel qualitative and quantitative validation methodology, that is based on a comparison between in silico results and postinterventional CT data, is presented. The validation methodology compares average stent diameters pseudo-continuously along the total length of the deployed SG. The validation of the in silico results shows very good agreement proving the potential of using in silico approaches in the preinterventional planning of EVAR. We consider models of bifurcated, marketed SGs as well as sophisticated models of patient-specific vessels that include intraluminal thrombus, calcifications and an anisotropic model for the vessel wall. We exemplarily show the additional benefit and applicability of in silico EVAR approaches to clinical cases by evaluating mechanical quantities with the potential to assess the quality of SG fixation and sealing such as contact tractions between SG and vessel as well as SG-induced tissue overstresses.
引用
收藏
页码:983 / 1004
页数:21
相关论文
共 374 条
[1]  
Acosta Santamaría V(2018)Model reduction methodology for computational simulations of endovascular repair Comput Methods Biomech Biomed Eng 21 1-10
[2]  
Daniel G(2015)Morphological and stent design risk factors to prevent migration phenomena for a thoracic aneurysm: a numerical analysis Med Eng Phys 37 23-33
[3]  
Perrin D(2011)Carotid artery stenting simulation: from patient-specific images to finite element analysis Med Eng Phys 33 281-289
[4]  
Albertini J(2013)Patient-specific aortic endografting simulation: from diagnosis to prediction Comput Biol Med 43 386-394
[5]  
Rosset E(2001)Results of an aortic endograft trial: impact of device failure beyond 12 months J Vasc Surg 33 55-63
[6]  
Avril S(2015)Towards efficient uncertainty quantification in complex and large-scale biomechanical problems based on a Bayesian multi-fidelity scheme Biomech Model Mechanobiol 14 489-513
[7]  
Altnji H-E(2012)CT artifacts: causes and reduction techniques Imaging Med 4 229-240
[8]  
Bou-Saïd B(2003)Predictive factors and clinical consequences of proximal aortic neck dilatation in 230 patients undergoing abdominal aorta aneurysm repair with self-expandable stent-grafts J Vasc Surg 37 1200-1205
[9]  
Walter-Le Berre H(2013)Ten-year results of endovascular abdominal aortic aneurysm repair from a large multicenter registry J Vasc Surg 58 324-332
[10]  
Auricchio F(1997)Aneurysm pressure following endovascular exclusion Eur J Vasc Endovasc Surg 13 85-87