A technique for calibrating a high dose rate 192Ir brachytherapy source

被引:0
作者
O. L. Fourie
T. G. Crabtree
机构
[1] Townsville Hospital,Medical Physics Department (IMB 4)
来源
Australasian Physical & Engineering Sciences in Medicine | 2012年 / 35卷
关键词
High dose rate brachytherapy; Reference air kerma rate; IAEA TECDOC 1274; IBA FC65-G ionisation chamber;
D O I
暂无
中图分类号
学科分类号
摘要
The reference air kerma rate of an 192Ir high dose rate brachytherapy source is determined based broadly on the International Atomic Energy Agency (IAEA) TECDOC 1274 code of practice. Since the primary standards dosimetry laboratory at the Australian Radiation Protection and Nuclear Safety Agency (ARPANSA) does not maintain a standard at 192Ir quality, the air kerma calibration coefficient of an IBA FC65-G Farmer type ionisation chamber is calculated using coefficients determined at 300 kV and 60Co qualities. The methodology proposed by Mainegra-Hing and Rogers [1] at 250 kV and 137Cs qualities is used. The validity of this approach is tested by performing Monte Carlo simulations to determine the chamber’s air kerma calibration coefficient at 192Ir quality. Very good agreement is obtained between values using these two methods. The reference air kerma rate is measured using the Farmer chamber in an in air jig. In addition the necessary correction factors are applied to the measured value. The reference air kerma rate determined in this way is compared to the value stated by the vendor of the 192Ir source on the source calibration certificate. Differences are with one exception less than 1%. It is concluded that because of the agreement between the values from the methodology used in this study and the source calibration certificate values this methodology can be used clinically.
引用
收藏
页码:85 / 92
页数:7
相关论文
共 53 条
[11]  
Reinstein E(1999)Correction factors for thick-walled ionization chambers in point source photon beams Med Phys 26 2441-2444
[12]  
Svensson GK(1994)Recombination factors for the cylindrical FC65-G ionization chamber in pulsed photon beams and the plane-parallel Roos ionization chamber in pulsed electron beams Nucl Instr Meth Phys Res A339 369-376
[13]  
Weller M(undefined)An EGSnrc investigation of cavity theory for ion chambers measuring air kerma undefined undefined undefined-undefined
[14]  
Wingfield L(undefined)Monte Carlo study of correction factors for Spencer–Attix cavity theory at photon energies at or above 100 keV undefined undefined undefined-undefined
[15]  
Rivard MJ(undefined)Spectra and air-kerma strength for encapsulated undefined undefined undefined-undefined
[16]  
Coursey BM(undefined)Ir sources undefined undefined undefined-undefined
[17]  
DeWerd LA(undefined)Activity determination of undefined undefined undefined-undefined
[18]  
Hanson WF(undefined)Ir solid sources by ionization chamber measurements using calculated corrections for self-absorption undefined undefined undefined-undefined
[19]  
Huq MS(undefined)undefined undefined undefined undefined-undefined
[20]  
Ibbott GS(undefined)undefined undefined undefined undefined-undefined