A technique for calibrating a high dose rate 192Ir brachytherapy source

被引:0
作者
O. L. Fourie
T. G. Crabtree
机构
[1] Townsville Hospital,Medical Physics Department (IMB 4)
来源
Australasian Physical & Engineering Sciences in Medicine | 2012年 / 35卷
关键词
High dose rate brachytherapy; Reference air kerma rate; IAEA TECDOC 1274; IBA FC65-G ionisation chamber;
D O I
暂无
中图分类号
学科分类号
摘要
The reference air kerma rate of an 192Ir high dose rate brachytherapy source is determined based broadly on the International Atomic Energy Agency (IAEA) TECDOC 1274 code of practice. Since the primary standards dosimetry laboratory at the Australian Radiation Protection and Nuclear Safety Agency (ARPANSA) does not maintain a standard at 192Ir quality, the air kerma calibration coefficient of an IBA FC65-G Farmer type ionisation chamber is calculated using coefficients determined at 300 kV and 60Co qualities. The methodology proposed by Mainegra-Hing and Rogers [1] at 250 kV and 137Cs qualities is used. The validity of this approach is tested by performing Monte Carlo simulations to determine the chamber’s air kerma calibration coefficient at 192Ir quality. Very good agreement is obtained between values using these two methods. The reference air kerma rate is measured using the Farmer chamber in an in air jig. In addition the necessary correction factors are applied to the measured value. The reference air kerma rate determined in this way is compared to the value stated by the vendor of the 192Ir source on the source calibration certificate. Differences are with one exception less than 1%. It is concluded that because of the agreement between the values from the methodology used in this study and the source calibration certificate values this methodology can be used clinically.
引用
收藏
页码:85 / 92
页数:7
相关论文
共 53 条
[1]  
Mainegra-Hing E(2006)On the accuracy of techniques for obtaining the calibration coefficient Med Phys 33 3340-3347
[2]  
Rogers DWO(1994) of Med Phys 21 581-618
[3]  
Kutcher GJ(2004)Ir HDR brachytherapy sources Med Phys 31 633-674
[4]  
Coia L(1991)Comprehensive QA for radiation oncology: Report of AAPM radiation therapy committee task group 40 Med Phys 18 462-467
[5]  
Gillin M(2005)Update of AAPM task group no. 43 report: a revised AAPM protocol for brachytherapy dose calculations Phys Med Biol 50 1961-1978
[6]  
Hanson WF(2008)Calibration of Australas Phys Eng Sci Med 31 332-338
[7]  
Leibel S(1990)Ir high-dose-rate afterloading systems Phys Med Biol 35 501-516
[8]  
Morton RJ(2004)Comparison of dosimetric standards of USA and France for HDR brachytherapy Phys Med Biol 49 5309-5318
[9]  
Palta JA(2003)Comparison of Med Phys 30 1211-1218
[10]  
Purdy L(2000)Ir air kerma calibration coefficients derived at ARPANSA using the interpolation method and at the National Physical Laboratory using a direct measurement Med Phys 27 1804-1813