On the optimization of low-cost FDM 3D printers for accurate replication of patient-specific abdominal aortic aneurysm geometry

被引:38
作者
Chung, Michael [1 ]
Radacsi, Norbert [1 ]
Robert, Colin [1 ]
McCarthy, Edward D. [1 ]
Callanan, Anthony [2 ]
Conlisk, Noel [2 ,3 ]
Hoskins, Peter R. [2 ,3 ]
Koutsos, Vasileios [1 ]
机构
[1] The School of Engineering, Institute for Materials and Processes, The University of Edinburgh, Robert Stevenson Road, Edinburgh
[2] The School of Engineering, Institute for Bioengineering, The University of Edinburgh, Max Born Crescent, Edinburgh
[3] Centre for Cardiovascular Sciences, The University of Edinburgh, 47 Little France Crescent, Edinburgh
关键词
3D printing; Abdominal aortic aneurysms; Accurate; Flexible; Rapid prototype; Thermoplastic polyurethane; Transparent;
D O I
10.1186/s41205-017-0023-2
中图分类号
学科分类号
摘要
Background: There is a potential for direct model manufacturing of abdominal aortic aneurysm (AAA) using 3D printing technique for generating flexible semi-transparent prototypes. A patient-specific AAA model was manufactured using fused deposition modelling (FDM) 3D printing technology. A flexible, semi-transparent thermoplastic polyurethane (TPU), called Cheetah Water (produced by Ninjatek, USA), was used as the flexible, transparent material for model manufacture with a hydrophilic support structure 3D printed with polyvinyl alcohol (PVA). Printing parameters were investigated to evaluate their effect on 3D–printing precision and transparency of the final model. ISO standard tear resistance tests were carried out on Ninjatek Cheetah specimens for a comparison of tear strength with silicone rubbers. Results: It was found that an increase in printing speed decreased printing accuracy, whilst using an infill percentage of 100% and printing nozzle temperature of 255 °C produced the most transparent results. The model had fair transparency, allowing external inspection of model inserts such as stent grafts, and good flexibility with an overall discrepancy between CAD and physical model average wall thicknesses of 0.05 mm (2.5% thicker than the CAD model). The tear resistance test found Ninjatek Cheetah TPU to have an average tear resistance of 83 kN/m, higher than any of the silicone rubbers used in previous AAA model manufacture. The model had lower cost (4.50 GBP per model), shorter manufacturing time (25 h 3 min) and an acceptable level of accuracy (2.61% error) compared to other methods. Conclusions: It was concluded that the model would be of use in endovascular aneurysm repair planning and education, particularly for practicing placement of hooked or barbed stents, due to the model’s balance of flexibility, transparency, robustness and cost-effectiveness. © 2018, The Author(s).
引用
收藏
相关论文
共 38 条
[1]  
Shimizu K., Mitchell R.N., Libby P., Inflammation and cellular immune responses in abdominal aortic aneurysms, Arterioscler Thromb Vasc Biol., 26, pp. 987-994, (2006)
[2]  
Vorp D.A., Biomechanics of abdominal aortic aneurysms, J Biomech, 40, pp. 1887-1902, (2009)
[3]  
Study Investigators TM. Aortic Wall Inflammation Predicts Abdominal Aortic Aneurysm Expansion, Rupture and Need for Surgical Repair. Circulation, (2017)
[4]  
Assar A.N., Zarins C.K., Ruptured abdominal aortic aneurysm: a surgical emergency with many clinical presentations, Postgrad. Med. J., 85, pp. 268-273, (2009)
[5]  
UK Small Aneurysm Trial Participants. Mortality results for randomised controlled trial of early elective surgery or ultrasonographic surveillance for small abdominal aortic aneurysms. Lancet, (1998)
[6]  
Lall P., Gloviczki P., Agarwal G., Duncan A.A., Kalra M., Hoskin T., Et al., Comparison of EVAR and open repair in patients with small abdominal aortic aneurysms: Can we predict results of the PIVOTAL trial? J. Vasc. Surg. [Internet]. The Society for, Vascular Surgery, 49, pp. 52-59, (2009)
[7]  
See K.W.M., Chui K.H., Chan W.H., Wong K.C., Chan Y.C., Evidence for endovascular simulation training: a systematic review. Eur. J. Vasc. Endovasc. Surg. [internet]. Elsevier Ltd
[8]  
2016
[9]  
51:441–51, Available from
[10]  
Saratzis A., Calderbank T., Sidloff D., Bown M.J., Davies R.S., Role of simulation in endovascular aneurysm repair (EVAR) training: a preliminary study. Eur. J. Vasc. Endovasc. Surg. [internet]. Elsevier Ltd