Failure of two commercial indexes and spectral parameters to reflect the pharmacodynamic effect of desflurane on EEG

被引:19
作者
Kreuer S. [1 ]
Bruhn J. [2 ]
Ellerkmann R. [3 ]
Ziegeler S. [1 ]
Kubulus D. [1 ]
Wilhelm W. [4 ]
机构
[1] Department of Anaesthesiology and Intensive Care Medicine, University of Saarland
[2] Department of Anaesthesiology, Nijmegen, UMC St. Radboud
[3] Department of Anaesthesiology and Intensive Care Medicine, University of Bonn, Bonn NRW
[4] Department of Anaesthesiology and Intensive Care Medicine, St.-Marien-Hospital, Luenen NRW
关键词
Bispectral index; Desflurane; Narcotrend; Pharmacodynamic model; Pharmacodynamic plateau;
D O I
10.1007/s10877-008-9116-1
中图分类号
学科分类号
摘要
Objective. We compared two PK/PD models, one with and one without a plateau effect. Bispectral (BIS, Aspect Medical Systems, Natick, MA, version XP) and Narcotrend® (NCT, MonitorTechnik, Bad Bramstedt, Germany, Version 4.0) indices were used as an electroencephalographic measure of desflurane drug effect. Methods. With IRB approval and informed consent we investigated 20 adult patients scheduled for radical prostatectomy. At least 45 minutes after induction of general anaesthesia, end-tidal concentrations of desflurane was varied between 3 and 10 vol%. To evaluate the relationship between concentrations and EEG indices, two different pharmacodynamic models were applied: A conventional model based on a single sigmoidal curve, and a novel model based on two sigmoidal curves for BIS and NCT values with and without burst suppression. The parameters of the models-were estimated by NONMEM V (GloboMax, Hanover, USA) by minimizing log likelihood. Statistical significance between the two models was calculated by the likelihood ratio test. Results. The maximum end-tidal desflurane concentration during the two concentrations ramps was 10.0±1.4 vol%. The mean BIS and NCT values decreased significantly but slightly with increasing end-tidal desflurane concentrations between 4 and 8 vol%. Therefore a two sigmoidal curves PK/PD model including a plateau describes the effects of desflurane on BIS and Narcotrend better than a single sigmoidal curve model. The difference between the log likelihood values of the new PK/PD model with two connected sigmoidal curves and the classical Emax model with one sigmoidal curve is 634 (P < 0.001) for the BIS monitor and 4089 (P < 0.001) for the NCT. Conclusions. BIS and Narcotrend are not useful to differentiate pharmacodynamic changes in the EEG between 4 and 9 vol% desflurane. © Springer Science+Business Media, LLC 2008.
引用
收藏
页码:149 / 158
页数:9
相关论文
共 21 条
[1]  
Johansen J.W., Sebel P.S., Development and clinical application of electroencephalographic bispectrum monitoring, Anesthesiology, 93, pp. 1336-1344, (2000)
[2]  
Rampil I.J., A primer for EEG signal processing in anesthesia, Anesthesiology, 89, pp. 980-1002, (1998)
[3]  
Schultz B., Kreuer S., Wilhelm W., Grouven U., Schultz A., The Narcotrend monitor. Development and interpretation algorithms, Anaesthesist, 52, pp. 1143-11148, (2003)
[4]  
Kreuer S., Biedler A., Larsen R., Altmann S., Wilhelm W., Narcotrend monitoring allows faster emergence and a reduction of drug consumption in propofol-remifentanil anesthesia, Anesthesiology, 99, pp. 34-41, (2003)
[5]  
Kreuer S., Bruhn J., Larsen R., Bialas P., Wilhelm W., Comparability of Narcotrend index and bispectral index during propofol anesthesia, Br J Anaesth, 93, pp. 235-240, (2004)
[6]  
Wilhelm W., Kreuer S., Larsen R., Narcotrend EEG monitoring during total intravenous anaesthesia in 4.630 patients, Anaesthesist, 51, pp. 980-988, (2002)
[7]  
Kreuer S., Bruhn J., Larsen R., Grundmann U., Wilhelm W., Application of Narcotrend index and bispectral index to the measurement of the EEG effects of isoflurane with and without burst suppression, Anesthesiology, 101, pp. 847-854, (2004)
[8]  
Kreuer S., Bruhn J., Walter E., Larsen R., Apfel C.C., Grundmann U., Biedler A., Wilhelm W., Comparative pharmacodynamic modeling using bispectral - And Narcotrend-index with and without a pharmacodynamic plateau during sevoflurane anesthesia, Anesth Analg, 106, 4, (2008)
[9]  
Hill A.V., The possible effect of the aggregration of the molecules of hemoglobin on its dissociation curves, J Physiol, 40, (1910)
[10]  
Shafer S.L., Pharmacokinetic and Pharmacodynamic Analysis With NONMEM-basic Concepts