Numerical investigations on the biomechanical compatibility of hip-joint endoprostheses

被引:0
|
作者
André Lutz
Udo Nackenhorst
机构
[1] Leibniz Universität Hannover,Institute of Mechanics and Computational Mechanics
来源
Archive of Applied Mechanics | 2010年 / 80卷
关键词
Stress adaptive bone remodeling; Hip-joint endoprosthetics; Finite element method;
D O I
暂无
中图分类号
学科分类号
摘要
A computational approach for studies on the biomechanical compatibility of hip implants based on detailed three-dimensional finite element models is presented. With the anticipation for a meaningful computational prediction of bone behavior under endoprosthetic treatment a general modeling approach taking into account the model generation from radiological data and the recovery of statically equivalent joint loads and muscle forces from a best fit between simulated and measured bone mass density distribution was developed. Computational results for three different endoprostheses systems are presented and compared with clinical studies. The potential for aiding the development process of new prosthesis designs regarding their biomechanical compatibility is demonstrated.
引用
收藏
页码:503 / 512
页数:9
相关论文
共 50 条
  • [1] Numerical investigations on the biomechanical compatibility of hip-joint endoprostheses
    Lutz, Andre
    Nackenhorst, Udo
    ARCHIVE OF APPLIED MECHANICS, 2010, 80 (05) : 503 - 512
  • [2] BIOMECHANICAL ASPECTS OF LOOSENING OF HIP-JOINT ENDOPROSTHESES
    PLITZ, W
    ORTHOPADE, 1989, 18 (05): : 344 - 349
  • [3] MOTILITY OF HIP-JOINT ENDOPROSTHESES
    WEBER, U
    HOSCHEK, J
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1979, 95 (1-2) : 95 - 104
  • [4] ON THE CEMENTLESS ANCHORAGE OF HIP-JOINT ENDOPROSTHESES
    QUACK, G
    RISCHKE, B
    BENSMANN, G
    KRAHL, H
    MARONNA, U
    SINGEWALD, M
    ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1991, 129 (05): : 453 - 459
  • [5] NUCLEAR IMAGING IN LOOSENING OF HIP-JOINT ENDOPROSTHESES
    ZILKENS, KW
    WICKE, A
    ZILKENS, J
    BULL, U
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1988, 107 (05) : 288 - 292
  • [6] REASONS FOR AND TACTICS OF REOPERATIONS OF TOTAL ENDOPROSTHESES OF THE HIP-JOINT
    CECH, O
    SOSNA, A
    SBORNIK LEKARSKY, 1984, 86 (2-3): : 70 - &
  • [7] CHARACTERISTIC BIOMECHANICAL VALUES FOR HIP-JOINT
    EHLER, E
    DIETZLAURSONN, M
    GATKE, C
    CHRISTMANN, C
    ANATOMISCHER ANZEIGER, 1977, 142 (03): : 237 - 242
  • [8] PREPARATION OF HIP-JOINT ENDOPROSTHESES FOR QUALITATIVE AND QUANTITATIVE MORPHOLOGICAL ANALYSIS
    HAHN, M
    DONATH, K
    DELLING, G
    CHIRURG, 1987, 58 (11): : 744 - 748
  • [9] ARTIFICIAL HIP-JOINT AND ITS BIOMECHANICAL DESTRUCTION
    GEBAUER, D
    ORLEY, H
    BLUMEL, G
    ARTIFICIAL ORGANS, 1979, 3 (03) : 297 - 297
  • [10] ARTHROSIS OF HIP-JOINT - RADIOLOGICAL INVESTIGATIONS
    CORTEREAL, M
    MUNCHENER MEDIZINISCHE WOCHENSCHRIFT, 1978, 120 (25): : 867 - 868