A new low-dose CT examination compared with standard-dose CT in the diagnosis of acute sinusitis

被引:0
作者
T. Hagtvedt
T. Aaløkken
J. Nøtthellen
A. Kolbenstvedt
机构
[1] Department of Radiology,
[2] Rikshospitalet,undefined
[3] 0027 Oslo,undefined
[4] Norway,undefined
来源
European Radiology | 2003年 / 13卷
关键词
Sinusitis CT Radiation dose;
D O I
暂无
中图分类号
学科分类号
摘要
A low-dose CT of the paranasal sinuses was designed with few, thin sections, non-uniform intersection gaps, low milliampere settings and avoidance of direct radiation to the eye lens. The low-dose CT was prospectively compared with standard-dose CT in patients with suspicion of acute sinusitis. Forty-seven patients were examined with low-dose CT immediately after standard-dose CT. The effective dose and the lens dose were calculated and compared. Using standard-dose CT as a gold standard the sensitivity and specificity of low-dose CT was calculated for each sinus group. The effective dose and the lens dose of the low-dose CT were reduced to, respectively, 3 and 2% of the standard-dose CT. The diagnostic yield of the low-dose CT with regard to acute sinusitis was good with a high specificity (≥96%) for all sinus groups. The sensitivity was also high (≥95%) except for the frontal sinus where the sensitivity was 83%. Low-dose CT offers considerable dose reduction and should be the standard for imaging patients with suspected acute inflammatory paranasal disease.