Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose

被引:0
作者
Sebastian Leschka
Paul Stolzmann
Florian T. Schmid
Hans Scheffel
Bjoern Stinn
Borut Marincek
Hatem Alkadhi
Simon Wildermuth
机构
[1] Kantonsspital St. Gallen,Institute of Radiology
[2] University Hospital Zurich,Institute of Diagnostic Radiology
来源
European Radiology | 2008年 / 18卷
关键词
Dual-source CT coronary angiography; Low kilovoltage; Radiation exposure; Contrast-to-noise ratio;
D O I
暂无
中图分类号
学科分类号
摘要
The purpose of this study was to investigate the effect of low kilovoltage dual-source computed tomography coronary angiography (CTCA) on qualitative and quantitative image quality parameters and radiation dose. Dual-source CTCA with retrospective ECG gating was performed in 80 consecutive patients of normal weight. Forty were examined with a standard protocol (120 kV/330mAs), 20 were examined at 100 kV/330mAs, and 20 at 100 kV/220mAs. Two blinded observers independently assessed image quality of each coronary segment and measured the image parameters noise, attenuation, and contrast-to-noise ratio (CNR). The effective radiation dose was calculated using CT dose volume index and the dose-length product. Diagnostic image quality was obtained in 99% of all coronary segments (1,127/1,140) without significant differences among the protocols. Image noise, attenuation, and CNR were significantly higher for 100 kV/330mAs (26 ± 3 HU, 549 ± 62 HU, 25.5 ± 3.2; each P < 0.01) and 100 kV/220mAs (27 ± 2 HU, 560 ± 43 HU, 25.0 ± 2.2; each P < 0.01) when compared to the 120-kV protocol (21 ± 2 HU, 317 ± 28 HU, 20.6 ± 1.7). There was no significant difference between the two 100-kV protocols. Estimated effective radiation dose of the 120-kV protocol (8.9 ± 1.2 mSv) was significantly higher than the 100 kV/330mAs (6.7 ± 0.8 mSv, P < 0.01) or 100 kV/220mAs (4.4 ± 0.6 mSv, P < 0.001) protocols. Dual-source CTCA with 100 kV is feasible in patients of normal weight, results in a diagnostic image quality with a higher CNR, and at the same time significantly reduces the radiation dose.
引用
收藏
页码:1809 / 1817
页数:8
相关论文
共 96 条
[1]  
Leschka S(2005)Accuracy of MSCT coronary angiography with 64-slice technology: first experience Eur Heart J 26 1482-1487
[2]  
Alkadhi H(2005)High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography Circulation 112 2318-2323
[3]  
Plass A(2005)Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography J Am Coll Cardiol 46 552-557
[4]  
Mollet NR(2006)Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control Eur Radiol 16 2739-2747
[5]  
Cademartiri F(2006)First performance evaluation of a dual-source CT (DSCT) system Eur Radiol 16 256-268
[6]  
van Mieghem CA(2007)Diagnostic accuracy of dual-source multi-slice CT-coronary angiography in patients with an intermediate pretest likelihood for coronary artery disease Eur Heart J. 28 2354-2360
[7]  
Raff GL(2007)Dual-source CT coronary angiography: image quality, mean heart rate, and heart rate variability AJR Am J Roentgenol 189 567-573
[8]  
Gallagher MJ(2006)Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates Circulation 113 1305-1310
[9]  
O’Neill WW(2008)Radiation dose estimates in dual-source computed tomography coronary angiography Eur Radiol. 18 592-599
[10]  
Goldstein JA(2000)Technique factors and image quality as functions of patient weight at abdominal CT Radiology 217 430-435