A case example and literature review of catastrophic wear before catastrophic failure: identification of trunnionosis and metallosis in metal-on-polyethylene hip arthroplasty prior to frank failure or fracture

被引:0
作者
Allison M. Hunter
Alan Hsu
Ashish Shah
Sameer M. Naranje
机构
[1] University of Alabama at Birmingham,Department of Orthopedic Surgery
来源
European Journal of Orthopaedic Surgery & Traumatology | 2019年 / 29卷
关键词
Arthroplasty; Metallosis; Metal-on-polyethylene; Metal-on-metal; Trunnionosis;
D O I
暂无
中图分类号
学科分类号
摘要
Though rare, there are documented failures of femoral prosthesis due to corrosion of the head–neck interface in total hip arthroplasty (THA), a phenomenon known as trunnionosis. This wear can result in metallosis, whereby metal debris scatters the surrounding soft tissues. We present on a 58-year-old female who presented with increase in hip and back pain 10 years following right THA using a metal-on-polyethylene construct with a large femoral head (44 mm). Aspiration withdrew metallic fluid, and intraoperative findings showed corrosion of the head–neck taper with surrounding metallosis and pseudocapsule formation. Despite advances in THA design, corrosion and wear between components still exists and may be cause for failure. We present on both the subtle clinical findings and the recommended workup when suspicion is high for trunnionosis, metallosis, or wear, ideally with identification prior to catastrophic failure such as component dislocation or fracture as previously reported.
引用
收藏
页码:711 / 715
页数:4
相关论文
共 7 条
  • [1] A case example and literature review of catastrophic wear before catastrophic failure: identification of trunnionosis and metallosis in metal-on-polyethylene hip arthroplasty prior to frank failure or fracture
    Hunter, Allison M.
    Hsu, Alan
    Shah, Ashish
    Naranje, Sameer M.
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2019, 29 (03) : 711 - 715
  • [2] The Result of Revision Total Hip Arthroplasty in Patients with Metallosis Following a Catastrophic Failure of a Polyethylene Liner
    Kwak, Hong Suk
    Yoo, Jeong Joon
    Lee, Young-Kyun
    Koo, Kyung-Hoi
    Yoon, Kang Sup
    Kim, Hee Joong
    CLINICS IN ORTHOPEDIC SURGERY, 2015, 7 (01) : 46 - 53
  • [3] A Case Report and Literature Review to Aid in the Management of Trunnion Failure in Hip Arthroplasty Patients: Can Trunnionosis and Prosthetic Joint Infection Co-Exist?
    Jagadale, Vivek
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2019, 11 (08)
  • [4] Gross trunnion failure in metal on polyethylene total hip arthroplasty—a systematic review of literature
    Tungish Bansal
    Sameer Aggarwal
    Mandeep Singh Dhillon
    Sandeep Patel
    International Orthopaedics, 2020, 44 : 609 - 621
  • [5] Gross trunnion failure in metal on polyethylene total hip arthroplasty-a systematic review of literature
    Bansal, Tungish
    Aggarwal, Sameer
    Dhillon, Mandeep Singh
    Patel, Sandeep
    INTERNATIONAL ORTHOPAEDICS, 2020, 44 (04) : 609 - 621
  • [6] Catastrophic failure due to massive osteolysis of both acetabular and femoral components in a metal-on-metal hip arthroplasty: A demonstrative case report
    Gutman G.
    Hershkovich O.
    Amit Y.
    Israeli A.
    European Journal of Orthopaedic Surgery & Traumatology, 2013, 23 (Suppl 2) : S225 - S228
  • [7] Unusual Case of a Trunnion Fracture Following a Revision Hip Arthroplasty Surgery: A Brief Review of the Literature and Discussion About Causes of Failure
    Vick, Hamilton P.
    Domingue, Grayson A.
    Richburg, Kacy
    Dean, Ryan
    Khan, Fahmida
    Thakral, Rishi
    ARTHROPLASTY TODAY, 2022, 18 : 103 - 106