Dual-source CT coronary angiography without heart rate or rhythm control in comparison with conventional coronary angiography

被引:0
作者
Xiang Ming Fang
Hong Wei Chen
Xiao Yun Hu
Jian Bao
Yin Chen
Zhen Yu Yang
Orla Buckley
Xiao Qing Wu
机构
[1] Wuxi People’s Hospital,Department of Radiology
[2] Nanjing Medical University,Department of Cardiology
[3] Wuxi People’s Hospital,Department of Radiology
[4] Nanjing Medical University,undefined
[5] Brigham and Women’s Hospital,undefined
来源
The International Journal of Cardiovascular Imaging | 2010年 / 26卷
关键词
Dual-source computed tomography; Angiography; Coronary artery disease;
D O I
暂无
中图分类号
学科分类号
摘要
To explore the value of dual-source CT angiography (DSCTA) in diagnosing coronary artery stenosis (CAS) without heart rate or rhythm control. Eighty-nine patients with different heart rates/rhythms underwent both DSCTA and conventional coronary angiography (CCA) in 1 week. The diagnostic quality of normal coronary arteries and stenosed segments using DSCTA and CCA were analyzed, respectively, with CCA as the gold standard. Kappa test was used to assess the intermodality agreement between DSCTA and CCA in grading CAS. The value of DSCTA in diagnosing CAS (>50% stenoses) were analyzed as well. The total evaluable rate of DSCTA in detecting coronary arteries was 98.8%. No significant difference between evaluable rates from different groups of heart rate (χ2 = 1.745, P > 0.05) was found. Sensitivity, specificity, positive and negative predictive values of DSCTA in detecting CAS and >50% stenoses were 97.9, 96.8, 89.5, 99.4 and 97.2, 95.5, 80.3, 99.4%, respectively; The inter-modality agreement between DSCTA and CCA in grading CAS was found to be excellent (k = 0.856, P < 0.01). DSCTA provides high accuracy and reliability for evaluation of CAS in a high suspected patient group without heart rate/rhythm control. It can be used as a powerful primary tool for the detection of CAS and a potential substitute of CCA.
引用
收藏
页码:323 / 331
页数:8
相关论文
共 63 条
[1]  
Achenbach S(2001)Detection of coronary artery stenoses by contrast enhanced retrospectively electrocardiographically-gated multislice spiral computed tomography Circulation 103 2535-2538
[2]  
Giesler T(2002)Non-invasive coronary angiography with high resolution multidetector-row computed tomography Eur Heart J 23 1714-1725
[3]  
Roper T(2003)Detection of coronary artery stenoses with thin slice multi-detector row spiral computed tomography and multiplanar reconstruction Circulation 107 664-666
[4]  
Kopp F(2004)Accuracy of thin-slice computed tomography in the detection of coronary stenoses Eur Heart J 25 1043-1048
[5]  
Schroeder S(2006)Noninvasive coronary angiography with 64-section CT: effect of average heart rate and heart rate variability on image quality Radiology 241 378-385
[6]  
Kuettner A(2006)Quantitative parameters of image quality in 64-slice computed tomography angiography of the coronary arteries Eur J Radiol 57 373-379
[7]  
Ropers D(2006)Image quality, motion artifacts, and reconstruction timing of 64-slice coronary computed tomography angiography with 0.33-second rotation speed Invest Radiol 41 436-442
[8]  
Baum U(2005)Accuracy of MSCT coronary angiography with 64-slice technology: first experience Eur Heart J 26 1482-1487
[9]  
Pohle K(2005)High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography Circulation 112 2318-2323
[10]  
Martuscelli E(2009)Assessment of image quality of 64-row dual source versus single source CT coronary angiography on heart rate: a phantom study Eur J Radiol 70 61-68