A Concentric Tube Robot System for Rigid Bronchoscopy: A Feasibility Study on Central Airway Obstruction Removal

被引:0
作者
Joshua B. Gafford
Scott Webster
Neal Dillon
Evan Blum
Richard Hendrick
Fabien Maldonado
Erin A. Gillaspie
Otis B. Rickman
S. Duke Herrell
Robert J. Webster
机构
[1] Vanderbilt University,Mechanical Engineering Department
[2] Virtuoso Surgical,undefined
[3] Inc.,undefined
[4] Vanderbilt University Medical Center,undefined
[5] Vanderbilt Institute for Surgery and Engineering (VISE),undefined
来源
Annals of Biomedical Engineering | 2020年 / 48卷
关键词
Central Airway Obstruction; Bronchoscopy; Robotics;
D O I
暂无
中图分类号
学科分类号
摘要
New robotic systems have recently emerged to assist with peripheral lung access, but a robotic system for rigid bronchoscopy has yet to be developed. We describe a new robotic system that can deliver thin robotic manipulators through the ports of standard rigid bronchoscopes. The manipulators bend and elongate to provide maneuverability of surgical tools at the endoscope tip, without endoscope motion. We describe an initial feasibility study on the use of this system to bronchoscopically treat a central airway obstruction (CAO). CAO is prevalent and can be life-threatening in patients with large tumors, and conventional rigid bronchoscopic treatments place patients at risk of complications including broken teeth, neck trauma and damage to oropharyngeal structures due to significant forces induced by bronchoscope tilting and manipulation. In this study, we used an ex vivo ovine airway model to demonstrate the ability of a physician using the robotic system to efficiently remove tissue and restore the airway. Pre- and post-operative CT scans showed that the robot was able to reduce the degree of airway obstruction stenosis from 75 to 14% on average for five CAO resections performed in an ex vivo animal model. Using cadaver experiments, we demonstrated the potential of the robotic system to substantially reduce the intraoperative forces applied to the patient’s head and neck (from 80.6 to 4.1 N). These preliminary results illustrate that CAO removal is feasible with our new rigid bronchoscopy robot system, and that this approach has the potential to reduce forces applied to the patient due to bronchoscope angulation, and thereby reduce the risk of complications encountered during CAO surgery.
引用
收藏
页码:181 / 191
页数:10
相关论文
共 218 条
[1]  
Breatnach E(1984)Dimensions of the normal human trachea Am. J. Roentgenol. 142 903-906
[2]  
Abbott G(1999)Multidisciplinary approach to management of postintubation tracheal stenoses Eur. Respir. J. 13 888-893
[3]  
Fraser R(2018)The future of robotic surgery Robotics 100 4-13
[4]  
Brichet A(2013)Safety and effectiveness of microdebrider bronchoscopy for the management of central airway obstruction Respirology 18 1011-1015
[5]  
Verkindre C(2011)Malignant airway obstruction: treating central airway obstruction in the oncologic setting UWOMJ 80 7-9
[6]  
Dupont J(2017)Early results of a safety and feasibility clinical trial of a novel single – port flexible robot for transoral robotic surgery Eur. Arch. Otorhinolaryngol. 274 3993-3996
[7]  
Carlier M(2011)Powered debridement of suprastomal granulation tissue to facilitate pediatric tracheotomy decannulation Int. J. Pediatr. Otorhinolaryngol. 75 1558-1561
[8]  
Darras J(1998)Malignant airway obstruction: recognition and management J. Emerg. Med. 16 83-92
[9]  
Wurtz A(2006)Outcome of treated advanced non-small cell lung cancer with and without central airway obstruction Chest 130 1803-1807
[10]  
Ramon P(2004)Central airway obstruction Am. J. Respir. Crit. Care Med. 169 1278-1297