Cochlear Implant Reliability: Reporting of Device Failures

被引:0
|
作者
Graham O’Neill
Neil S. Tolley
机构
[1] St Mary’s Hospital,Dept of Otolaryngology, Head and Neck Surgery
来源
Indian Journal of Otolaryngology and Head & Neck Surgery | 2020年 / 72卷
关键词
Implanted device; Reliability; Cumulative failure percentage; Cumulative failure rate; Instantaneous failure rate;
D O I
暂无
中图分类号
学科分类号
摘要
The aim of this study was to investigate the adequacy of the reporting of cochlear implant device failures. Data from a parallel study involving over 6300 children [1] was used to calculate the instantaneous failure rate for explantations. We found that this is comparable to what manufacturers term ‘Cumulative Failure Percentage’ (CFP). This finding raises concerns about the information provided by manufacturers on the reliability of their implants.
引用
收藏
页码:326 / 328
页数:2
相关论文
共 50 条
  • [1] Cochlear Implant Reliability: Reporting of Device Failures
    O'Neill, Graham
    Tolley, Neil S.
    INDIAN JOURNAL OF OTOLARYNGOLOGY AND HEAD & NECK SURGERY, 2020, 72 (03) : 326 - 328
  • [2] Cochlear Implant Reliability: On the Reporting of Rates of Revision Surgery
    Graham O’Neill
    Neil S. Tolley
    Indian Journal of Otolaryngology and Head & Neck Surgery, 2020, 72 : 257 - 266
  • [3] Cochlear Implant Reliability: On the Reporting of Rates of Revision Surgery
    O'Neill, Graham
    Tolley, Neil S.
    INDIAN JOURNAL OF OTOLARYNGOLOGY AND HEAD & NECK SURGERY, 2020, 72 (02) : 257 - 266
  • [4] Validity and reliability of the Nijmegen Cochlear Implant Questionnaire in German
    Plath, Michaela
    Sand, Matthias
    van de Weyer, Philipp S.
    Baierl, Kilian
    Praetorius, Mark
    Plinkert, Peter K.
    Baumann, Ingo
    Zaoui, Karim
    HNO, 2022, 70 (06) : 422 - 435
  • [5] International Classification of Reliability for Implanted Cochlear Implant Receiver Stimulators
    Battmer, Rolf-Dieter
    Backous, Douglas D.
    Balkany, Thomas J.
    Briggs, Robert J. S.
    Gantz, Bruce J.
    van Hasselt, Andrew
    Kim, Chong Sun
    Kubo, Takeshi
    Lenarz, Thomas
    Pillsbury, Harold C., III
    O'Donoghue, Gerard M.
    OTOLOGY & NEUROTOLOGY, 2010, 31 (08) : 1190 - 1193
  • [6] Reliability of connectivity for the assessment of hearing perception in cochlear implant recipients
    Brotto, Davide
    Catelan, Dolores
    Pegolo, Margherita
    Marioni, Gino
    Galoforo, Nicole
    Sorrentino, Flavia
    Biggeri, Annibale
    Lovato, Elisa
    De Filippis, Cosimo
    Trevisi, Patrizia
    INTERNATIONAL JOURNAL OF AUDIOLOGY, 2024,
  • [7] Multicenter evaluation of Neurelec Digisonic® SP cochlear implant reliability
    Radulescu, Luminita
    Cozma, Sebastian
    Niemczyk, Casimir
    Guevara, Nicolas
    Gahide, Ivan
    Economides, John
    Lavieille, Jean-Pierre
    Meller, Renaud
    Bebear, Jean-Pierre
    Radafy, Emilien
    Bordure, Philippe
    Djennaoui, Djamel
    Truy, Eric
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2013, 270 (04) : 1507 - 1512
  • [8] Multicenter evaluation of Neurelec Digisonic® SP cochlear implant reliability
    Luminiţa Rădulescu
    Sebastian Cozma
    Casimir Niemczyk
    Nicolas Guevara
    Ivan Gahide
    John Economides
    Jean-Pierre Lavieille
    Renaud Meller
    Jean-Pierre Bébéar
    Emilien Radafy
    Philippe Bordure
    Djamel Djennaoui
    Eric Truy
    European Archives of Oto-Rhino-Laryngology, 2013, 270 : 1507 - 1512
  • [9] Cochlear implant failures and reimplantation: A 30-year analysis and literature review
    Lane, Ciaran
    Zimmerman, Kim
    Agrawal, Sumit
    Parnes, Lorne
    LARYNGOSCOPE, 2020, 130 (03) : 782 - 789
  • [10] Assessing competence in cochlear implant surgery using the newly developed Cochlear Implant Surgery Assessment Tool
    Frendo, Martin
    Frithioff, Andreas
    Konge, Lars
    Foghsgaard, Soren
    Mikkelsen, Peter Trier
    Sorensen, Mads Solvsten
    Caye-Thomasen, Per
    Andersen, Steven Arild Wuyts
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2022, 279 (01) : 127 - 136