CD34+ cell collection efficiency does not correlate with the pre-leukapheresis hematocrit

被引:0
|
作者
J Mehta
Y Oyama
J Winter
S Williams
M Tallman
S Singhal
M Villa
T Shook
R Burt
A Traynor
G Soff
S Masarik
G Ramsey
L Gordon
机构
[1] The Hematopoietic Stem Cell Transplant Program,Division of Hematology/Oncology, Department of Internal Medicine
[2] Northwestern University Medical School and The Robert H Lurie Comprehensive Cancer Center of Northwestern University,Division of Immunotherapy and Autoimmune Diseases, Department of Internal Medicine
[3] Northwestern University Medical School,undefined
来源
Bone Marrow Transplantation | 2001年 / 28卷
关键词
apheresis; CD34; cells; cell separator; collection efficiency, FAHCT; hematocrit; leukapheresis;
D O I
暂无
中图分类号
学科分类号
摘要
One hundred and seventy-seven large-volume leukapheresis procedures performed on 91 patients over a 15 month period were reviewed to see if the pre-apheresis hematocrit (Hct) affected the CD34+ cell collection efficiency (CE) of the Fenwal CS 3000 Plus cell separator. The Hct was 0.174–0.461 (median 0.317), and the peripheral blood CD34+ cell count 2–2487 per μl (median 21). The total CD34+ cell quantity collected was 3.0–2677.2 × 106 (median 113.0). Based on the number of CD34+cells contained in the blood volume processed (23.3–37303.2 × 106; median 318.0), the CE was 1.7–87.5% (median 30.3). No correlation was found between the Hct and CE (r2 = 0.0034; P = 0.44) or the total CD34+ cell quantity collected (r2 = 0.0040; P = 0.40). CEs for Hct <0.25 (median CE 36%), Hct 0.25–0.299 (median CE 30%) and Hct 0.30 (median CE 30%) were comparable. As expected, highly significant correlations were seen between the CD34+ cell quantities collected and quantities processed (r2 = 0.59; P < 10−6) as well as the peripheral blood CD34+ cell counts (r2 = 0.60; P < 10−6). We conclude that the minimum acceptable Hct or hemoglobin level for leukapheresis should be dictated by clinical circumstances because it does not affect stem cell collection. Bone Marrow Transplantation (2001) 28, 597–601.
引用
收藏
页码:597 / 601
页数:4
相关论文
共 50 条
  • [1] CD34+ cell collection efficiency does not correlate with the pre-leukapheresis hematocrit
    Mehta, J
    Oyama, Y
    Winter, J
    Williams, S
    Tallman, M
    Singhal, S
    Villa, M
    Shook, T
    Burt, R
    Traynor, A
    Soff, G
    Masarik, S
    Ramsey, G
    Gordon, L
    BONE MARROW TRANSPLANTATION, 2001, 28 (06) : 597 - 601
  • [2] Validation of a formula for predicting daily CD34+ cell collection by leukapheresis
    Rosenbaum, Eric R.
    O'connell, Bernadette
    Cottler-Fox, Michele
    CYTOTHERAPY, 2012, 14 (04) : 461 - 466
  • [3] Can the stem cell mobilization technique influence CD34+ cell collection efficiency of leukapheresis procedures in patients with hematologic malignancies?
    A Gidron
    A Verma
    M Doyle
    L Boggio
    A Evens
    L Gordon
    S Singhal
    M Tallman
    S Williams
    J Winter
    J Mehta
    Bone Marrow Transplantation, 2005, 35 : 243 - 246
  • [4] Crude collection efficiency of CD34+ hematopoietic stem cell apheresis
    Castillo-Aleman, Yandy Marx
    JOURNAL OF CLINICAL APHERESIS, 2024, 39 (01)
  • [5] Can the stem cell mobilization technique influence CD34+cell collection efficiency of leukapheresis procedures in patients with hematologic malignancies?
    Gidron, A
    Verma, A
    Doyle, M
    Boggio, L
    Evens, A
    Gordon, L
    Singhal, S
    Tallman, M
    Williams, S
    Winter, J
    Mehta, J
    BONE MARROW TRANSPLANTATION, 2005, 35 (03) : 243 - 246
  • [6] CD34+ Collection Efficiency as a Function of Blood Volumes Processed in Pediatric Autologous Peripheral Blood Stem Cell Collection
    Dubrovsky, Leonid
    Wong, Edward C. C.
    Perez-Albuerne, Evelio
    Loechelt, Brett
    Kamani, Naynesh
    Sande, Jane
    Mintz, Kathy
    Paul, Wendy
    Luban, Naomi L. C.
    Rood, Brian R.
    Fry, Terry
    JOURNAL OF CLINICAL APHERESIS, 2011, 26 (03) : 131 - 137
  • [7] A new model for predicting on the basis of CD34+ the timing of leukapheresis cell and hematopoietic progenitor cell levels
    Teng, Hao-Wei
    Hsiao, Liang-Tsai
    Chaou, Shu-Chou
    Gau, Jyh-Pyng
    Lee, Tzu-Chi
    Shih, Ying-Yih
    Liu, Chun-Yu
    Hong, Ying-Chung
    Chen, Ming-Huang
    Chang, Mu-Hsin
    Yang, Ya-Hsu
    Chen, Po-Min
    JOURNAL OF CLINICAL APHERESIS, 2007, 22 (04) : 195 - 203
  • [8] Influence of preapheresis clinical factors on the efficiency of CD34+ cell collection by large-volume apheresis
    C Sarkodee-Adoo
    I Taran
    C Guo
    F Buadi
    R Murthy
    E Cox
    R Lopez
    S Westphal
    S Shope
    B O'Connell
    L Wethers
    B Meisenberg
    Bone Marrow Transplantation, 2003, 31 : 851 - 855
  • [9] Comparison of CD34+ cell collection efficiency on the COBE Spectra and Fenwal CS-3000 Plus
    Ford, CD
    Lehman, C
    Strupp, A
    Kelley, L
    JOURNAL OF CLINICAL APHERESIS, 2002, 17 (01) : 17 - 20
  • [10] Absolute numbers of peripheral blood CD34+hematopoietic stem cells prior to a leukapheresis procedure as a parameter predicting the efficiency of stem cell collection
    Galtseva, I. V.
    Davydova, Yu. O.
    Gaponova, T. V.
    Kapranov, N. M.
    Kuzmina, L. A.
    Troitskaya, V. V.
    Gribanova, E. O.
    Kravchenko, S. K.
    Mangasarova, Ya. K.
    Zvonkov, E. E.
    Parovichnikova, E. N.
    Mendeleeva, L. P.
    Savchenko, V. G.
    TERAPEVTICHESKII ARKHIV, 2017, 89 (07) : 18 - 24