Realization of Negative Gauge Pressure Standard in the National Metrology Institute of Indonesia

被引:0
作者
Adindra Vickar Ega
Gigin Ginanjar
R. Rudi Anggoro Samodro
机构
[1] National Standardization Agency of Indonesia (BSN),
来源
MAPAN | 2020年 / 35卷
关键词
Negative pressure; Pressure balance; Bell jar; Absolute pressure mode; Atmospheric pressure; Barometer;
D O I
暂无
中图分类号
学科分类号
摘要
Accurate negative gauge pressure measurement is the key factor for many scientific and industrial applications such as clean room technology, medical treatment, and pharmaceutical production. However, there are not much publications regarding to the technical guide when it comes to negative pressure measurement, although several comparison of negative gauge pressure has been performed in Europe. In this study, two methods for negative pressure calibration have been successfully implemented by National Standardization Agency of Indonesia with using pressure balance. The first method is by generating the negative pressure in the bell jar of the pressure balance, and the second method is by using pressure balance in absolute pressure mode and additional barometric pressure measuring instrument. Both methods measurement results show linear relation from the negative to positive pressure and show good agreement by each other with the normalized error (En) less than 0.40 from – 1000 to 1000 hPa with uncertainty up to 0.073 hPa. The expanded uncertainty of the second method is relatively larger due to dominant contribution from the uncertainty of the barometer used. Therefore, it can be concluded that both methods can be used in order to provide traceability for negative gauge pressure measurement instrument in Indonesia.
引用
收藏
页码:97 / 103
页数:6
相关论文
共 38 条
[1]  
Tsao TH(2010)The Effect of Doorway Size on the Negative Pressure in Airborne Infection Isolation Room J. Occup. Saf. Health 18 307-319
[2]  
Chang CP(2016)Negative pressure wound therapy in orthopaedic surgery Orthopaedics & Traumatology Surgery & Research 29 035007-156
[3]  
Chuang CY(2018)Calibration Methods for Negative Gauge Pressure Down to − 100 kPa Meas. Sci. Technol. 53 07010-595
[4]  
Juang YJ(2016)EURAMET.M.P-S9: comparison in the negative gauge pressure range − 950 to 0 hPa Metrologia 47 7007–07007-undefined
[5]  
Lim SJ(2010)Supplementary Comparison: Negative gauge pressure comparison: range -95 kPa to +95 kPa (EURAMET Project 1131) Metrologia 51 07010-undefined
[6]  
Dai YT(2014)Final Report on EURAMET.M.P-S12—Bilateral supplementary comparison of the national pressure standards of CMI and INRIM in the range 300 Pa to 15 kPa of negative gauge pressure Metrologia 43 151-undefined
[7]  
Robert N(2009)The evaluation of key comparison data Primary Etalonnage of Negative Gauge Pressures Using Pressure Balances at the Czech Metrology Institute, Materiali in technologije/Materials and technology 39 589-undefined
[8]  
Bentouati D(2002)undefined Metrologia undefined undefined-undefined
[9]  
Durgut Y(undefined)undefined undefined undefined undefined-undefined
[10]  
Otal P(undefined)undefined undefined undefined undefined-undefined