Anatomical and physiological variables influencing measurement of regional cerebral oxygen saturation by near infrared spectroscopy using the Sensmart Model X-100TM

被引:0
作者
Yuichi Tanaka
Manzo Suzuki
Kenji Yoshitani
Atsuhiro Sakamoto
Hiroyasu Bito
机构
[1] Nippon Medical School,Department of Anesthesiology, Musashikosugi Hospital
[2] National Cardiovascular Center,Department of Anesthesiology
[3] Nippon Medical School,Department of Anesthesiology
来源
Journal of Clinical Monitoring and Computing | 2021年 / 35卷
关键词
Neuro monitoring; Regional oxygen saturation; Effect of anatomical variation; Sensmart model X-100; NIRO-200X;
D O I
暂无
中图分类号
学科分类号
摘要
The Sensmart Model X-100 (Nonin Medical Inc, Plymouth, MN, USA) is a relatively new device that possesses two sets of emitters and detectors and uses near infrared spectroscopy (NIRS) to measure regional cerebral oxygen saturation (rSO2). The value of rSO2 obtained by other NIRS devices is affected by physiological and anatomical variables such as hemoglobin concentration, area of cerebrospinal fluid (CSF) layer and skull thickness. The effects of these variables have not yet been determined in measurement of rSO2 by Sensmart Model X-100. We examined the effects of area of CSF, hemoglobin concentration, and skull thickness on the values of rSO2 measured by Sensmart Model X-100 and tissue oxygen index (TOI) measured by NIRO-200NX (Hamamatsu Photonix, Hamamatsu, Japan). Forty neurosurgical, cardiac and vascular surgical patients who underwent preoperative computed tomographic (CT) scan of the brain were enrolled in this study. Regional cerebral oxygen saturation (rSO2) at the forehead was measured sequentially by NIRO-200NX and by Sensmart Model X-100. Simultaneously, mean arterial pressure, hemoglobin concentration, and partial pressure of carbon dioxide in arterial blood (PaCO2) were measured. To evaluate the effects of anatomical factors on rSO2, we measured skull thickness and area of CSF layer using CT images of the brain. Multiple regression analysis was used to examine the relationships between the rSO2 values and anatomical and physiological factors. The area of the CSF layer and hemoglobin concentration had significant associations with rSO2 measured by the Sensmart Model X-100, whereas none of the studied variables was significantly associated with TOI. The measurement of rSO2 by Sensmart Model X-100 is not affected by the skull thickness of patients. Area of the CSF layer and hemoglobin concentration may be the main biases in measurement of rSO2 by Sensmart Model X-100.
引用
收藏
页码:1063 / 1068
页数:5
相关论文
共 101 条
[1]  
Terakado T(2019)Effectiveness of near-infrared spectroscopy (NIRO-200NX, Pulse mode) for risk management in carotid artery stenting World Neurosurg 131 e425-432
[2]  
Marushima A(2012)Cerebral oximetry during cardiac surgery: the association between cerebral oxygen saturation and perioperative patient variables J Cardiothorac Vasc Anesth 26 1015-1021
[3]  
Koyama Y(2007)Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study Anesth Analg 104 51-58
[4]  
Tsuruta W(2000)Evaluation of a cerebral oximeter as a monitor of cerebral ischemia during carotid endarterectomy Anesthesiology 93 964-970
[5]  
Takigawa T(2007)A retrospective study of changes in cerebral oxygenation using a cerebral oximeter in older patients undergoing prolonged major abdominal surgery Eur J Anaesthesiol 24 230-234
[6]  
Ito Y(2018)Benchmarking of four near-infrared spectroscopy devices for long term use in neonates Klin Padistr 230 240-244
[7]  
Hino T(2007)Effects of hemoglobin concentration, skull thickness, and area of cerebrospinal fluid layer on near-infrared spectroscopy measurements Anesthesiology 106 458-462
[8]  
Sato M(2012)Development and validation of a cerebral oximeter. Capable of absolute accuracy J Cardiothorac Vasc Anesth 26 1007-1014
[9]  
Hayakawa M(2012)Impact of extracranial contamination on regional cerebral oxygen saturation. A comparison of three cerebral oximetry technologies Anesthesiology 116 834-840
[10]  
Ishikawa E(2009)Variability in thickness of human skull bones and sternum—an autopsy experience J Forensic Med Toxicol 26 26-31