Ultraviolet photolysis of urine for suppression of color quenching prior to liquid scintillation counting of tritium

被引:0
|
作者
Yoko Watanabe
Jun Kuwabara
机构
[1] Japan Atomic Energy Agency,Radiation Protection Department, Nuclear Science Research Institute
来源
关键词
UV photolysis; Liquid scintillation counter; Tritium; Color quenching; Rapid analysis;
D O I
暂无
中图分类号
学科分类号
摘要
In order to reduce the color quenching in the measurement of tritium in urine by liquid scintillation counting , UV irradiation was applied to decompose the organic substances in the sample. Urine was decolorized under UV irradiation in the presence of hydrogen peroxide. As a result, color quenching was considerably suppressed and higher counting efficiency of tritium was obtained. This UV treatment made it possible to increase the urine content in the sample from 2 to 40% (v/v) without significant decrease of counting efficiency. Either higher sensitivity or shorter analysis time was achieved in the tritium measurement by the augmentation of urine content. When the measurement time was 30 min, the detection limit of tritium defined as 3s was 0.03 Bq/ml. At the expense of some sensitivity (set at a detection limit of 0.3 Bq/ml), the measurement time was shortened to 0.5 min. These results will make a great improvement to routine tritium monitoring as well as to emergency monitoring in mass tritium exposure.
引用
收藏
页码:547 / 550
页数:3
相关论文