Fracture of a ceramic liner in a total hip arthroplasty with a sandwich cup

被引:0
作者
Dragos Popescu
X. Gallart
S. Garcia
G. Bori
X. Tomas
J. Riba
机构
[1] University of Barcelona,Department of Orthopedics and Trauma Surgery (Hip Unit), Hospital Clínic
[2] University of Barcelona,Department of Radiology, Hospital Clínic
来源
Archives of Orthopaedic and Trauma Surgery | 2008年 / 128卷
关键词
Hip arthroplasty; Liner fracture; Complication;
D O I
暂无
中图分类号
学科分类号
摘要
Alumina ceramic-on-ceramic total hip arthroplasty (THA) has been widely used due to its advantages such as low wear, scratch resistance, wettable surface and relatively low biological reactivity of the wear particles. Nevertheless, this material in THA still persists to be one of the major concerns about the risk of fracture, due to its brittleness. Many authors have reported a fracture of the ceramic head but few reported a fracture of the ceramic acetabular insert. In order to reduce the rigidity of the ceramic-on-ceramic coupling and prevent an impingement between the rim of the ceramic liner and the metal neck of the femoral stem, a modular acetabular component with a sandwich insertion (alumina/polyethylene/titanium) was proposed. We report the fracture of the ceramic acetabular liner of such a ceramic sandwich cup due to a slightly retroverted position of the cup that causes an impingement between the femoral stem and the rim of the insert. The fracture occurred 3 years after the operation without trauma. At revision the entire cup was replaced using a polyethylene liner without inner ceramic liner.
引用
收藏
页码:783 / 785
页数:2
相关论文
共 30 条
[1]  
Arenas A(1999)Ceramic femoral head fractures in total hip replacement Int Orthop 23 351-352
[2]  
Tejero A(2004)Concerns about ceramics in THA Clin Orthop Relat Res 429 73-79
[3]  
Garbayo A(2007)Fracture of the ABC Clin Orthop Relat 453 97-102
[4]  
Arias C(2007) ceramic liner: a report of three cases Clin Orthop Relat Res 458 106-110
[5]  
Barbadillo J(2003)Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty J Arthroplasty 18 658-661
[6]  
Lecumberri J(1997)Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup J Bone Joint Surg Am 79A 118-121
[7]  
Barrack RL(2006)Fracture of a ceramic femoral head after a revision operation. A case report J Arthroplasty 21 1072-1073
[8]  
Burak C(1992)Fracture of a ceramic femoral head after ceramic-on-ceramic total hip arthroplasty Clin Orthop 282 31-46
[9]  
Skinner HB(undefined)On the material and the tribology of alumina–alumina couplings for hip joint prostheses undefined undefined undefined-undefined
[10]  
Diwanji SR(undefined)undefined undefined undefined undefined-undefined