Neutron and gamma ray doses from a 252Cf brachytherapy source in a water phantom

被引:0
作者
Ibahim Al-Saihati
A. A. Naqvi
机构
[1] King Fahd University of Petroleum and Minerals,Department of Electrical Engineering
[2] King Fahd University of Petroleum and Minerals,Department of Physics
来源
Journal of Radioanalytical and Nuclear Chemistry | 2013年 / 296卷
关键词
Brachytherapy; Californium-252; Monte Carlo Simulation; Neutron dose; Gamma dose;
D O I
暂无
中图分类号
学科分类号
摘要
Axial and radial doses of Neutrons and gamma rays from an Isotron 252Cf Brachytherapy source were calculated in a Water phantom using Maienshein’s prompt fission gamma rays data and Maxwellian neutron energy spectrum. It was observed that neutron dose due to the source casing thickness does not contribute significantly to the total dose. Further the calculated secondary gamma ray dose rate is very small compared to the calculated primary gamma dose rate. Neutron and secondary gamma ray dose calculated in this study agree with the published data. Results of this study will be presented here.
引用
收藏
页码:963 / 966
页数:3
相关论文
共 10 条
[1]  
Tacev T(2004)undefined Nuclear instruments and methods in physics research, section b 213 629-undefined
[2]  
Strnad V(2001)undefined Klinicka Onkologie 14 59-undefined
[3]  
Ptackova B(1999)undefined Med Phys 26 1503-undefined
[4]  
Tacev T(undefined)undefined undefined undefined undefined-undefined
[5]  
Ptackova B(undefined)undefined undefined undefined undefined-undefined
[6]  
Strnad V(undefined)undefined undefined undefined undefined-undefined
[7]  
Koukalova H(undefined)undefined undefined undefined undefined-undefined
[8]  
Krystof V(undefined)undefined undefined undefined undefined-undefined
[9]  
Rasovska O(undefined)undefined undefined undefined undefined-undefined
[10]  
Rivard MJ(undefined)undefined undefined undefined undefined-undefined