摘要
The performance of the piezo-composite transducer is greatly influenced by the matching layers. Based on finite element analysis and artificial intelligence, an optimization design method is proposed to optimize the matching layers of a 2–2 piezo-composite transducer for biomedical imaging. The neural networks are trained by the finite element analysis data to build up the mapping relationship between the thickness of matching layers and the performance. The optimization criteria are established based on the performance (centre frequency, bandwidth, and peak-to-peak voltage) and minimizing the material consumption. The thicknesses of matching layers are optimized by using a particle swarm optimization algorithm. The optimized thicknesses of both matching layers are 49 μm. The optimized centre frequency, bandwidth, and peak-to-peak voltage are 9.1 MHz, 72.5%, and 2.26 V, which can nearly achieve the designed targets (9.5 MHz, 75%, and 2.20 V). According to the optimized thicknesses of matching layers, the fabricated 2–2 piezo-composite transducer exhibits a centre frequency of 9.8 MHz, a bandwidth of 79.6%, and a peak-to-peak voltage of 2.11 V verifies the effectiveness and availability of the proposed method. Then, the piezo-composite transducer is hard-pressed for ultrasonic imaging. The press-focused piezo-composite transducer has an insertion loss of -19.1 dB at 10 MHz, a lateral resolution of 125 μm at 4 mm, and an axial resolution of 132 μm. The good resolution was confirmed by scanning the pig eyeball, which suggests that the 2–2 piezo-composite transducer with optimized matching layers has great potential in biomedical imaging.
引用
收藏
相关论文
共 50 条
  • [31] Analytical model and optimal design of a d33-mode active layer for the lightweight unimorph piezo-composite actuator
    Ngoc-Trung Nguyen
    Yoon, Bum-Soo
    Park, Ki-Hoon
    Yoon, Kwang-Joon
    JOURNAL OF ELECTROCERAMICS, 2011, 26 (1-4) : 175 - 184
  • [32] Analytical model and optimal design of a d33-mode active layer for the lightweight unimorph piezo-composite actuator
    Ngoc-Trung Nguyen
    Bum-Soo Yoon
    Ki-Hoon Park
    Kwang-Joon Yoon
    Journal of Electroceramics, 2011, 26 : 175 - 184
  • [33] ULTRASONIC TRANSDUCER DESIGN FOR UNIFORM INSONATION IN BIOMEDICAL ULTRASOUND
    HARRISON, GH
    BALCERKUBICZEK, EK
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1986, 33 (01) : 97 - 97
  • [34] Actuation performance improvement of lightweight piezo-composite actuator with an inter digitated electrode actuation layer
    Tran, Anh Kim
    Yoon, K. J.
    Goo, Nam Seo
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2007, 51 : S16 - S21
  • [35] A high temperature and broadband immersion 1-3 piezo-composite transducer for accurate inspection in harsh environments
    Xue, Q
    Stanton, M
    Elfbaum, G
    2003 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2003, : 1372 - 1375
  • [36] Design and demonstration of a biomimetic wing section using a lightweight piezo-composite actuator (LIPCA)
    Lim, SM
    Lee, S
    Park, HC
    Yoon, KJ
    Goo, NS
    SMART MATERIALS AND STRUCTURES, 2005, 14 (04) : 496 - 503
  • [37] Optimization Design of Ultrasonic Transducer With Multimatching Layer
    Li, Zhaoxi
    Chen, Dongdong
    Fei, Chunlong
    Li, Di
    Feng, Wei
    Yang, Yintang
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2021, 68 (06) : 2202 - 2211
  • [38] 40 MHz piezo-composite linear array for medical imaging and integration in a high resolution system
    Bantignies, Claire
    Mauchamp, Pascal
    Dufait, Remi
    Levassort, Franck
    Mateo, Tony
    Gregoire, Jean-Marc
    Ossant, Frederic
    2011 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2011, : 226 - 229
  • [39] An improved design of Piezo-Composite Actuator used as the artificial muscle for bio-inspired robots
    Nguyen, Ngoc-Trung
    Yoon, Bum-Soo
    Yoon, Kwang-Joon
    2007 IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND BIOMIMETICS, VOLS 1-5, 2007, : 7 - 12
  • [40] Enhancement of Ultrasonic Transducer Bandwidth by Acoustic Impedance Gradient Matching Layer
    Zhu, Ke
    Ma, Jinpeng
    Qi, Xudong
    Shen, Bingzhong
    Liu, Yang
    Sun, Enwei
    Zhang, Rui
    SENSORS, 2022, 22 (20)