Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT

被引:0
|
作者
Medhat M. Osman
Christian Cohade
Yuji Nakamoto
Richard L. Wahl
机构
[1] The Johns Hopkins University School of Medicine, Department of Radiology, Division of Nuclear Medicine
[2] Saint Louis University, Department of Internal Medicine, Division of Nuclear Medicine
来源
European Journal of Nuclear Medicine and Molecular Imaging | 2003年 / 30卷
关键词
PET/CT; Image fusion; Artifact; Respiratory motion;
D O I
暂无
中图分类号
学科分类号
摘要
PET-CT scanners allow generation of transmission maps from CT. The use of CT attenuation correction (CTAC) instead of germanium-68 attenuation correction (Ge AC) might be expected to cause artifacts on reconstructed emission images if differences in respiratory status exist between the two methods of attenuation correction. The aim of this study was to evaluate for possible respiratory motion artifacts (RMA) in PET images attenuation corrected with CT from PET-CT in clinical patients. PET-CT scans were performed using a Discovery LS PET-CT system in 50 consecutive patients (23 males, 27 females; mean age 58.2 years) with known or suspected malignancy. Both CTAC and Ge AC transmission data obtained during free tidal breathing were used to correct PET emission images. Cold artifacts at the interface of the lungs and diaphragm, believed to be due to respiratory motion (RMA), that were seen on CTAC images but not on the Ge AC images were evaluated qualitatively on a four-point scale (0, no artifact; 1, mild artifact; 2, moderate artifact; 3, severe artifact). RMA was also measured for height. Curvilinear cold artifacts paralleling the dome of the diaphragm at the lung/diaphragm interface were noted on 84% of PET-CT image acquisitions and were not seen on the 68Ge-corrected images; however, these artifacts were infrequently severe. In conclusion, RMA of varying magnitude were noted in most of our patients as a curvilinear cold area at the lung/diaphragm interface, but were not diagnostically problematic in these patients.
引用
收藏
页码:603 / 606
页数:3
相关论文
共 50 条
  • [1] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT
    Osman, MM
    Cohade, C
    Nakamoto, Y
    Wahl, RL
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (04) : 603 - 606
  • [2] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT.
    Osman, MM
    Cohade, C
    Wahl, RL
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 305P - 305P
  • [3] Reduction of attenuation correction artifacts in PET-CT
    Nuyts, Johan
    Stroobants, Sigrid
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1895 - 1899
  • [4] Correction for oral contrast artifacts in CT attenuation-corrected PET images obtained by combined PET/CT
    Nehmeh, SA
    Erdi, YE
    Kalaigian, H
    Kolbert, KS
    Pan, TS
    Yeung, H
    Squire, O
    Sinha, A
    Larson, SM
    Humm, JL
    JOURNAL OF NUCLEAR MEDICINE, 2003, 44 (12) : 1940 - 1944
  • [5] Deep learning-aided respiratory motion compensation in PET/CT: addressing motion induced resolution loss, attenuation correction artifacts and PET-CT misalignment
    Lu, Yihuan
    Kang, Fei
    Zhang, Duo
    Li, Yue
    Liu, Hao
    Sun, Chen
    Zeng, Hao
    Shi, Lei
    Zhao, Yumo
    Wang, Jing
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2024, 52 (01) : 62 - 73
  • [6] Influence of attenuation correction (AC) in positron emission tomography (PET) and combined PET-CT on artifacts of hip prostheses
    Goerres, GW
    Burger, CN
    Berthold, T
    Buck, A
    RADIOLOGY, 2001, 221 : 386 - 386
  • [7] Respiratory motion artefacts in pet imaging with CT attenuation correction
    Vale, J.
    Costa, D.
    Fonseca, G.
    Patrina, J.
    Faria, D.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S272 - S272
  • [8] Respiratory motion artefacts in PET imaging with CT attenuation correction
    Vale, J.
    Costa, D. C.
    Fonseca, G.
    Patrina, J.
    Faria, D.
    Ponte, F.
    Teixeira, J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 : S355 - S355
  • [9] Respiratory Motion Correction Strategies in Thoracic PET-CT Imaging
    Nehmeh, Sadek A.
    PET CLINICS, 2013, 8 (01) : 29 - +
  • [10] Lung Motion Correction of Respiratory Gated 3D PET-CT Images
    Dawood, M.
    Jiang, X.
    Lang, N.
    Schober, O.
    Schaefers, K. P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 : S95 - S95