Detection of Breast Cancer with a Computer-Aided Detection Applied to Full-Field Digital Mammography

被引:0
|
作者
Ryusuke Murakami
Shinichiro Kumita
Hitomi Tani
Tamiko Yoshida
Kenichi Sugizaki
Tomoyuki Kuwako
Tomonari Kiriyama
Kenta Hakozaki
Emi Okazaki
Keiko Yanagihara
Shinya Iida
Shunsuke Haga
Shinichi Tsuchiya
机构
[1] Graduate School of Medicine,Department of Radiology
[2] Nippon Medical School,Department Surgery for Organ Function and Biological Regulation
[3] Graduate School of Medicine,Division of Diagnostic Pathology
[4] Nippon Medical School,undefined
[5] Nippon Medical School Hospital,undefined
来源
Journal of Digital Imaging | 2013年 / 26卷
关键词
Breast neoplasm; Cancer detection; Computer-aided diagnosis (CAD); Digital mammography;
D O I
暂无
中图分类号
学科分类号
摘要
A study was conducted to evaluate the sensitivity of computer-aided detection (CAD) with full-field digital mammography in detection of breast cancer, based on mammographic appearance and histopathology. Retrospectively, CAD sensitivity was assessed in total group of 152 cases for subgroups based on breast density, mammographic presentation, lesion size, and results of histopathological examination. The overall sensitivity of CAD was 91 % (139 of 152 cases). CAD detected 100 % (47/47) of cancers manifested as microcalcifications; 98 % (62/63) of those manifested as non-calcified masses; 100 % (15/15) of those manifested as mixed masses and microcalcifications; 75 % (12/16) of those manifested as architectural distortions, and 69 % (18/26) of those manifested as focal asymmetry. CAD sensitivity was 83 % (10/12) for cancers measuring 1–10 mm, 92 % (37/40) for those measuring 11–20 mm, and 92 % (92/100) for those measuring >20 mm. There was no significant difference in CAD detection efficiency between cancers in dense breasts (88 %; 69/78) and those in non-dense breasts (95 %; 70/74). CAD showed a high sensitivity of 91 % (139/152) for the mammographic appearance of cancer and 100 % sensitivity for identifying cancers manifested as microcalcifications. Sensitivity was not influenced by breast density or lesion size. CAD should be effective for helping radiologists detect breast cancer at an earlier stage.
引用
收藏
页码:768 / 773
页数:5
相关论文
共 50 条
  • [31] Retrospective analysis of a computer-aided detection (CAD) system in full-field digital mammography in correlation to tumor histology
    Obenauer, S
    Sohns, C
    Werner, C
    Grabbe, E
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2005, 177 (08): : 1103 - 1109
  • [32] Who could benefit the most from using a computer-aided detection system in full-field digital mammography?
    Na Young Jung
    Bong Joo Kang
    Hyeon Sook Kim
    Eun Suk Cha
    Jae Hee Lee
    Chang Suk Park
    In Young Whang
    Sung Hun Kim
    Yeong Yi An
    Jae Jeong Choi
    World Journal of Surgical Oncology, 12
  • [33] Who could benefit the most from using a computer-aided detection system in full-field digital mammography?
    Jung, Na Young
    Kang, Bong Joo
    Kim, Hyeon Sook
    Cha, Eun Suk
    Lee, Jae Hee
    Park, Chang Suk
    Whang, In Young
    Kim, Sung Hun
    An, Yeong Yi
    Choi, Jae Jeong
    WORLD JOURNAL OF SURGICAL ONCOLOGY, 2014, 12
  • [34] Computer-aided detection of masses in full-field digital mammography using screen-film mammograms for training
    Kallenberg, Michiel
    Karssemeijer, Nico
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (23): : 6879 - 6891
  • [35] Computer-aided detection of breast masses on full field digital mammogranris
    Wei, J
    Sahiner, B
    Hadjiiski, LM
    Chan, HP
    Petrick, N
    Helvie, MA
    Roubidoux, MA
    Ge, J
    Zhou, C
    MEDICAL PHYSICS, 2005, 32 (09) : 2827 - 2838
  • [36] Computer-aided diagnosis (CAD) of images obtained on full-field digital mammography
    Nishikawa, RM
    Giger, ML
    Yarusso, LM
    Kupinski, MA
    Baehr, AE
    Venta, LA
    RADIOLOGY, 1999, 213P : 229 - 229
  • [37] Computer-aided breast cancer detection in screening mammography
    Bick, U
    Giger, ML
    Schmidt, RA
    Nishikawa, RM
    Wolverton, DE
    Doi, K
    DIGITAL MAMMOGRAPHY '96, 1996, 1119 : 97 - 103
  • [38] Computer-Aided Detection and Diagnosis of Breast Abnormalities in Digital Mammography
    Bozek, Jelena
    Delac, Kresimir
    Grgic, Mislav
    PROCEEDINGS ELMAR-2008, VOLS 1 AND 2, 2008, : 45 - 52
  • [39] Microcalcification detection in full-field digital mammograms: A fully automated computer-aided system
    Basile, T. M. A.
    Fanizzi, A.
    Losurdo, L.
    Bellotti, R.
    Bottigli, U.
    Dentamaro, R.
    Didonna, V
    Fausto, A.
    Massafra, R.
    Moschetta, M.
    Tamborra, P.
    Tangaro, S.
    La Forgia, D.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 64 : 1 - 9
  • [40] Accuracy of breast cancer detection with full-field digital mammography (FFDM) and integral computer-aided detection (CAD) correlated with breast density as assessed by a new automated volumetric breast density measurement system
    Pinker, K.
    Milner, S.
    Stephen, D.
    Mokbel, K.
    Perry, N. M.