Scintillating fiber phantom dosimeter to measure real-time dose distribution of Co-60 teletherapy source

被引:0
作者
Hyun Young Shin
Sang Hun Shin
Si Won Song
Seunghyun Cho
Bongsoo Lee
机构
[1] Chung-Ang University,School of Energy Systems Engineering
[2] Soongsil University,Department of Organic Materials and Fiber Engineering
来源
Optical Review | 2019年 / 26卷
关键词
Scintillating fiber; Phantom dosimeter; CMOS sensor camera; Co-60 teletherapy source; Dose distribution;
D O I
暂无
中图分类号
学科分类号
摘要
In this study, we developed a scintillating fiber phantom dosimeter (SFPD) that can measure the transverse and longitudinal dose distributions of a Co-60 teletherapy source. The SFPD system is composed of organic scintillating fibers, polymethyl methacrylate (PMMA) phantoms, and a complementary metal–oxide–semiconductor (CMOS) sensor camera. When the phantom dosimeter is irradiated with gamma rays from a Co-60 teletherapy source, scintillating lights are emitted from the scintillating fibers inserted in the phantom dosimeter. The scintillating images are measured using a CMOS camera in real time and analyzed using an image processing program. The proposed phantom dosimeter has many advantages such as real-time measurement, easy calibration, and immunity to electromagnetic interference. In addition, it can measure complex dose distributions more precisely, because it has a high spatial resolution due to the small size of organic scintillating fibers. To evaluate the performance of the proposed phantom dosimeter, transverse and longitudinal dose distributions are measured using the CMOS sensor camera. Dose distributions are measured according to the field sizes and compared to the results obtained using an EBT3 film and MCNPX calculations.
引用
收藏
页码:391 / 395
页数:4
相关论文
共 31 条
[1]  
Rice RK(1987)Measurements of dose distributions in small beams of 6 MV X-rays Phys. Med. Biol. 32 1087-1099
[2]  
Hansen JL(2011)Calibration of GafChromic XR-RV3 radiochromic film for skin dose measurement using standardized X-ray spectra and a commercial flatbed scanner Med. Phys. 38 1919-1930
[3]  
Svensson GK(2015)Performance evaluation of a multichannel all-in-one phantom dosimeter for dose measurement of diagnostic X-ray beam Sensors 15 28490-28501
[4]  
Siddon RL(2006)Accurate skin dose measurements using radiochromic film in clinical applications Med. Phys. 33 1116-1124
[5]  
McCabe BP(2006)Water equivalent plastic scintillation detectors in radiation therapy Radiat. Prot. Dosim. 120 1-6
[6]  
Speidel MA(2000)Water equivalence of plastic organic scintillators in megavoltage radiotherapy bremsstrahlung beams Phys. Med. Biol. 45 1885-1895
[7]  
Pike TL(2006)Development and characterization of a tissue equivalent plastic scintillator based dosimetry system Med. Phys. 33 96-105
[8]  
Van Lysel MS(undefined)undefined undefined undefined undefined-undefined
[9]  
Jeon H(undefined)undefined undefined undefined undefined-undefined
[10]  
Yoo WJ(undefined)undefined undefined undefined undefined-undefined